Suppr超能文献

经静脉除颤的导联系统优化

Lead system optimization for transvenous defibrillation.

作者信息

Gold M R, Foster A H, Shorofsky S R

机构信息

Department of Medicine, University of Maryland School of Medicine, Baltimore, USA.

出版信息

Am J Cardiol. 1997 Nov 1;80(9):1163-7. doi: 10.1016/s0002-9149(97)00633-4.

Abstract

Lead systems that include an active pectoral shell reduce defibrillation thresholds and permit transvenous defibrillation in nearly all patients. A further improvement in defibrillation efficacy is desirable to allow for smaller pulse generators with a reduced maximum output. Accordingly, the purpose of this study was to compare defibrillation thresholds with multiple transvenous lead systems including those with an active pectoral shell to determine which system would optimize defibrillation energy requirements. This prospective study was performed on 21 consecutive patients. Each subject was evaluated with 3 lead configurations with the order of testing randomized. The configurations were a dual coil transvenous lead (lead), the distal right ventricular coil and pectoral pulse generator shell (unipolar), and all 3 components (triad). The right ventricular coil was the cathode for the first phase of the biphasic defibrillation waveform. Delivered energy at defibrillation threshold was 11.2 +/- 3.4 J for the lead configuration, 10.1 +/- 5.2 J for the unipolar configuration, and 7.8 +/- 3.6 J for the triad configuration (p <0.01). Leading edge voltage (p <0.01) and shock impedance (p <0.001) were also decreased for the triad configuration compared with the lead or unipolar configurations, whereas peak current was minimized with the unipolar configuration (p <0.01). We conclude that the combination of a dual coil, transvenous lead and an active pectoral shell reduces defibrillation energy requirements compared with either the lead alone or unipolar configuration. Moreover, the defibrillation thresholds were < or =15 J in all patients using the triad lead system.

摘要

包含有源胸段外壳的导联系统可降低除颤阈值,并使几乎所有患者都能进行经静脉除颤。为了使用最大输出功率降低的更小脉冲发生器,需要进一步提高除颤效果。因此,本研究的目的是比较多种经静脉导联系统(包括带有有源胸段外壳的系统)的除颤阈值,以确定哪种系统能优化除颤能量需求。这项前瞻性研究对21例连续患者进行。每位受试者用3种导联配置进行评估,测试顺序随机。配置分别为双线圈经静脉导联(导联)、右心室远端线圈和胸段脉冲发生器外壳(单极)以及所有3个组件(三联体)。右心室线圈是双相除颤波形第一阶段的阴极。导联配置在除颤阈值时的输送能量为11.2±3.4J,单极配置为10.1±5.2J,三联体配置为7.8±3.6J(p<0.01)。与导联或单极配置相比,三联体配置的前沿电压(p<0.01)和电击阻抗(p<0.001)也降低,而单极配置的峰值电流最小(p<0.01)。我们得出结论,与单独的导联或单极配置相比,双线圈经静脉导联和有源胸段外壳的组合可降低除颤能量需求。此外,使用三联体导联系统的所有患者的除颤阈值均≤15J。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验