Suppr超能文献

有源胸壁脉冲发生器对心房除颤阈值无益处。

Lack of benefit of an active pectoral pulse generator on atrial defibrillation thresholds.

作者信息

Gold Michael R, Shorofsky Stephen R, Bouhouch Rachida, Joy Robert A, Kirby D Scott, Osman Ahmed F, Rashba Eric J

机构信息

Division of Cardiology, Medical University of South Carolina, Charleston 29425, USA.

出版信息

J Cardiovasc Electrophysiol. 2002 Apr;13(4):332-5. doi: 10.1046/j.1540-8167.2002.00332.x.

Abstract

INTRODUCTION

Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator leads, which has led to the development of combined atrial and ventricular devices. For ventricular defibrillation, use of an active pectoral electrode (active can) in the shocking pathway markedly reduces defibrillation thresholds (DFTs). However, the effect of an active pectoral can on atrial defibrillation is unknown.

METHODS AND RESULTS

This study was a prospective, randomized, paired comparison of two shock configurations on atrial DFTs in 33 patients. The lead system evaluated was a dual-coil transvenous defibrillation lead with a left pectoral pulse generator emulator. Shocks were delivered either between the right ventricular coil and proximal atrial coil (lead) or between the right ventricular coil and an active can in common with the atrial coil (active can). Delivered energy at DFT was 4.2 +/- 4.1 J in the lead configuration and 5.0 +/- 3.7 J in the active can configuration (P = NS). Peak current was 32% higher with an active can (P < 0.01), whereas shock impedance was 18% lower (P < 0.001). Moreover, a low threshold (< or = 3 J) was observed in 61% of subjects in the lead configuration but in only 36% in the active can configuration (P < 0.05). There were no clinical predictors of the atrial DFT.

CONCLUSION

These results indicate that low atrial DFTs can be achieved using a transvenous ventricular defibrillation lead. Because no benefit was observed with the use of an active pectoral electrode for atrial defibrillation, programmable shock vectors may be useful for dual-chamber implantable cardioverter defibrillators.

摘要

引言

使用标准的植入式心脏复律除颤器电极可实现心房除颤,这推动了心房和心室联合装置的发展。对于心室除颤,在电击传导路径中使用主动式胸壁电极(主动式罐电极)可显著降低除颤阈值(DFT)。然而,主动式胸壁罐电极对心房除颤的影响尚不清楚。

方法与结果

本研究是一项前瞻性、随机、配对比较33例患者两种电击配置下心房DFT的研究。评估的导联系统是带有左胸壁脉冲发生器模拟器的双线圈经静脉除颤导联。电击分别在右心室线圈和近端心房线圈(导联)之间或右心室线圈和与心房线圈共用的主动式罐电极之间进行。导联配置下DFT时的释放能量为4.2±4.1J,主动式罐电极配置下为5.0±3.7J(P=无显著差异)。主动式罐电极的峰值电流高32%(P<0.01),而电击阻抗低18%(P<0.001)。此外,导联配置下61%的受试者观察到低阈值(≤3J),而主动式罐电极配置下仅为36%(P<0.05)。没有心房DFT的临床预测因素。

结论

这些结果表明,使用经静脉心室除颤导联可实现低心房DFT。由于未观察到使用主动式胸壁电极进行心房除颤有任何益处,可编程电击向量可能对双腔植入式心脏复律除颤器有用。

相似文献

1
Lack of benefit of an active pectoral pulse generator on atrial defibrillation thresholds.
J Cardiovasc Electrophysiol. 2002 Apr;13(4):332-5. doi: 10.1046/j.1540-8167.2002.00332.x.
2
Optimization of atrial defibrillation with a dual-coil, active pectoral lead system.
J Cardiovasc Electrophysiol. 2004 Jul;15(7):790-4. doi: 10.1046/j.1540-8167.2004.03684.x.
3
Clinical predictors of atrial defibrillation thresholds with a dual-coil, active pectoral lead system.
Heart Rhythm. 2005 Jan;2(1):49-54. doi: 10.1016/j.hrthm.2004.10.033.
4
Atrial defibrillation with a transvenous lead: a randomized comparison of active can shocking pathways.
J Am Coll Cardiol. 1999 Aug;34(2):358-62. doi: 10.1016/s0735-1097(99)00214-4.
5
Coronary sinus electrode does not reduce atrial defibrillation thresholds.
Heart Rhythm. 2006 Jun;3(6):647-52. doi: 10.1016/j.hrthm.2006.02.1029. Epub 2006 Mar 3.
6
Optimization of transvenous coil position for active can defibrillation thresholds.
J Cardiovasc Electrophysiol. 2000 Jan;11(1):25-9. doi: 10.1111/j.1540-8167.2000.tb00731.x.
7
Effect of an active abdominal pulse generator on defibrillation thresholds with a dual-coil, transvenous ICD lead system.
J Cardiovasc Electrophysiol. 2006 Jun;17(6):617-20. doi: 10.1111/j.1540-8167.2006.00374.x.
8
Lead system optimization for transvenous defibrillation.
Am J Cardiol. 1997 Nov 1;80(9):1163-7. doi: 10.1016/s0002-9149(97)00633-4.
9
The effect of shock configuration and delivered energy on defibrillation impedance.
Pacing Clin Electrophysiol. 1999 Jan;22(1 Pt 2):165-8. doi: 10.1111/j.1540-8159.1999.tb00325.x.
10
Atrial defibrillation thresholds of electrode configurations available to an atrioventricular defibrillator.
J Cardiovasc Electrophysiol. 2001 Aug;12(8):957-64. doi: 10.1046/j.1540-8167.2001.00957.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验