Gold Michael R, Shorofsky Stephen R, Bouhouch Rachida, Joy Robert A, Kirby D Scott, Osman Ahmed F, Rashba Eric J
Division of Cardiology, Medical University of South Carolina, Charleston 29425, USA.
J Cardiovasc Electrophysiol. 2002 Apr;13(4):332-5. doi: 10.1046/j.1540-8167.2002.00332.x.
Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator leads, which has led to the development of combined atrial and ventricular devices. For ventricular defibrillation, use of an active pectoral electrode (active can) in the shocking pathway markedly reduces defibrillation thresholds (DFTs). However, the effect of an active pectoral can on atrial defibrillation is unknown.
This study was a prospective, randomized, paired comparison of two shock configurations on atrial DFTs in 33 patients. The lead system evaluated was a dual-coil transvenous defibrillation lead with a left pectoral pulse generator emulator. Shocks were delivered either between the right ventricular coil and proximal atrial coil (lead) or between the right ventricular coil and an active can in common with the atrial coil (active can). Delivered energy at DFT was 4.2 +/- 4.1 J in the lead configuration and 5.0 +/- 3.7 J in the active can configuration (P = NS). Peak current was 32% higher with an active can (P < 0.01), whereas shock impedance was 18% lower (P < 0.001). Moreover, a low threshold (< or = 3 J) was observed in 61% of subjects in the lead configuration but in only 36% in the active can configuration (P < 0.05). There were no clinical predictors of the atrial DFT.
These results indicate that low atrial DFTs can be achieved using a transvenous ventricular defibrillation lead. Because no benefit was observed with the use of an active pectoral electrode for atrial defibrillation, programmable shock vectors may be useful for dual-chamber implantable cardioverter defibrillators.
使用标准的植入式心脏复律除颤器电极可实现心房除颤,这推动了心房和心室联合装置的发展。对于心室除颤,在电击传导路径中使用主动式胸壁电极(主动式罐电极)可显著降低除颤阈值(DFT)。然而,主动式胸壁罐电极对心房除颤的影响尚不清楚。
本研究是一项前瞻性、随机、配对比较33例患者两种电击配置下心房DFT的研究。评估的导联系统是带有左胸壁脉冲发生器模拟器的双线圈经静脉除颤导联。电击分别在右心室线圈和近端心房线圈(导联)之间或右心室线圈和与心房线圈共用的主动式罐电极之间进行。导联配置下DFT时的释放能量为4.2±4.1J,主动式罐电极配置下为5.0±3.7J(P=无显著差异)。主动式罐电极的峰值电流高32%(P<0.01),而电击阻抗低18%(P<0.001)。此外,导联配置下61%的受试者观察到低阈值(≤3J),而主动式罐电极配置下仅为36%(P<0.05)。没有心房DFT的临床预测因素。
这些结果表明,使用经静脉心室除颤导联可实现低心房DFT。由于未观察到使用主动式胸壁电极进行心房除颤有任何益处,可编程电击向量可能对双腔植入式心脏复律除颤器有用。