Reproducibility of stress echocardiography using intravenous injection of ultrasound contrast agent (BY 963).

作者信息

Leischik R, Kuhlmann C, Bruch C, Jeremias A, Buck T, Erbel R

机构信息

University of Essen, Cardiology Department, Germany.

出版信息

Int J Card Imaging. 1997 Oct;13(5):387-94. doi: 10.1023/a:1005822920962.

Abstract

UNLABELLED

Despite the widespread use of stress echocardiography, its reproducibility is still limited by high interobserver variability. Therefore, the purpose of the present study was to improve the reproducibility of a stress (exercise) echocardiography using a new transpulmonary ultrasound agent (BY 963). Stress echocardiography was performed in 12 healthy volunteers with suboptimal endocardial border delineation during exercise echocardiography. A special 45 degrees lateral tilted bike stress echocardiography table was used for exercise testing. Echocardiographic images were recorded on-line at rest and during exercise on a video tape and additionally digitized on-line on a stress echo computer. End-diastolic (EDVml), end-systolic (ESVml) volume and ejection fraction (EF%) were estimated in the 4-chamber view. The measurements were performed before and after injection of 2.5 ml and 5 ml BY963 at rest and in maximal exercise. A new contrast agent (BY 963) leads to a sufficient contrast effect for the left ventricular cavity after intravenous administration and permits a good delineation of left the endocardial border. The interobserver variability was determined using blinded investigation by two observers. The correlation of EDV and ESV determination at rest was r = 0.68/0.33, after 2.5 ml BY 963 r = 0.97/0.93 and after 5 ml BY 963 r = 0.90/0.93. The correlation for EDV and ESV during exercise was r = 0.52/0.33, after 2.5 ml BY 963 r = 0.88/0.80 and after 5 ml BY 963 r = 0.95/0.92. At rest mean EF without contrast was 61 +/- 6%/67 +/- 7% (r = 0. 130), after 2.5 ml BY 963 i.v. 69 +/- 8%/72 +/- 7% (r = 0.82) and after 5 ml BY 963 i.v. 73 +/- 8%/73 +/- 8% (r = 0.98%) respectively. In exercise, mean EF without contrast was 68 +/- 8%/70 +/- 6 (r = 0.013), after 2.5 ml BY 963 83 +/- 6%/81 +/- 5 and after 5 ml 83 +/- 4%/82 +/- 3 (r = 0.86).

SUMMARY

The estimation of the end-systolic volume in exercise will be improved significantly and the estimated EF values will be higher compared to EF values obtained without contrast application. Transpulmonary contrast echocardiography for analysis of left ventricular volumes and ejection fraction can be routinely used in stress echocardiography. Intravenous administration of BY 963 improves the reproducibility of quantitative analysis of left ventricular function in healthy volunteers. Further studies in patients with cardiac diseases are required to corroborate this observation.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验