Suppr超能文献

使用新型超声雾化器进行机械通气(间歇正压通气)时的气溶胶产生和气溶胶液滴大小分布

Aerosol production and aerosol droplet size distribution during mechanical ventilation (IPPV) with a new ultrasonic nebulizer.

作者信息

Kemming G I, Kreyling W, Habler O, Merkel M, Kleen M, Welte M, Messmer K, Zwissler B

机构信息

Institute of Anaesthesiology, Ludwig-Maximilians Universiy Munich, Klinikum Grosshadern, Marchioninistr. 15, Munich D-81377, Germany.

出版信息

Eur J Med Res. 1996 Apr 18;1(7):321-7.

PMID:9364032
Abstract

Administration of drugs via the airway is increasingly practiced in ICU- and surgical patients. For this purpose, aerosols may be produced by either jet nebulization or ultrasonic droplet generation. In mechanically ventilated patients, aerosol delivery is often insufficient. The influence of the ventilatory pattern on nebulizer efficacy is poorly understood. In the present in vitro study we determined the efficacy of a new ultrasonic nebulizer in delivering aerosolized epoprostenol using defined ventilator settings. We determined aerosol delivery rates, the aerosol droplet size distribution and the impact of the connection tubing on drug delivery, applying adult and infant ventilation patterns. Aerosol production rates ranged from 0.28 to 0.57 ml per minute. Using an adult ventilator setting volume controlled ventilation (CMV) led to a higher aerosol production rate than pressure controlled ventilation (PCV) at identical tidal volumes and mean airway pressures (0.57 ml/min,CMV vs 0.39 ml/min, PCV). With an infant ventilator setting, nebulizer rates were lower than those found for the adult ventilator setting, but did not differ substantially between CMV and PCV mode (0.29 ml/min, CMV vs 0.28 ml/min, PCV). Aerosol delivery rates distal to the endotracheal tube changed according to aerosol production rates (adult mode: 0.18 ml/min, CMV vs 0.10 ml/min, PCV; infant mode: 0.03 ml/min, both CMV and PCV). In the infant ventilation mode, a higher percentage of the aerosol was trapped in the catheter mount as compared to the adult ventilation mode. Mass median droplet diameters for each of the four ventilator settings were almost identical (4.63 to 5.09 micron) and smaller than indicated in the product specifications (8 micron). Delivery rates and sizes of droplets delivered by the new ultrasonic nebulizer SUN 345(R) agree well with previously reported data from comparable settings using diverse nebulizer devices.

摘要

在重症监护病房(ICU)患者和外科手术患者中,通过气道给药的做法越来越普遍。为此,可通过喷射雾化或超声液滴生成来产生气溶胶。在机械通气患者中,气溶胶输送往往不足。通气模式对雾化器疗效的影响尚不清楚。在本体外研究中,我们使用确定的通气设置,测定了一种新型超声雾化器输送雾化依前列醇的疗效。我们测定了气溶胶输送速率、气溶胶液滴大小分布以及连接管对药物输送的影响,采用了成人和婴儿通气模式。气溶胶产生速率范围为每分钟0.28至0.57毫升。在相同潮气量和平均气道压力下,使用成人通气设置的容量控制通气(CMV)比压力控制通气(PCV)产生更高的气溶胶产生速率(0.57毫升/分钟,CMV对0.39毫升/分钟,PCV)。在婴儿通气设置下,雾化器速率低于成人通气设置下的速率,但在CMV和PCV模式之间没有显著差异(0.29毫升/分钟,CMV对0.28毫升/分钟,PCV)。气管导管远端的气溶胶输送速率根据气溶胶产生速率而变化(成人模式:0.18毫升/分钟,CMV对0.10毫升/分钟,PCV;婴儿模式:0.03毫升/分钟,CMV和PCV均如此)。与成人通气模式相比,在婴儿通气模式下,更高比例的气溶胶被困在导管接口中。四种通气设置下的质量中值液滴直径几乎相同(4.63至5.09微米),且小于产品规格中所示(8微米)。新型超声雾化器SUN 345(R)输送的液滴速率和大小与之前使用不同雾化器设备的类似设置所报告的数据非常吻合。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验