Rheineck-Leyssius A T, Kalkman C J
Department of Anaesthesia, Twenteborg Hospital, Almelo, The Netherlands.
Br J Anaesth. 1997 Oct;79(4):460-4. doi: 10.1093/bja/79.4.460.
In a prospective, randomized study, we have investigated the effects of two arbitrary pulse oximeter lower alarm limit (LAL) settings (90% = group 90, n = 320 and 85% = group 85, n = 327) on the incidence of hypoxaemia in the recovery room. In group 90, we calculated the theoretical effect of elimination of transient episodes of low pulse oximeter oxyhaemoglobin saturation (SpO2) by introducing a time delay between the onset of the alarm condition and triggering of the alarm. When only hypoxaemic episodes lasting more than 1 min were included, SpO2 < or = 90% occurred in 11% of patients in group 90 and in 20% in group 85 (relative risk (RR) 1.84, confidence interval (CI) 1.26-2.69; P < 0.01). Hypoxaemia < or = 85% occurred in 2% of patients in group 90 and in 6% in group 85 (RR 3.10, CI 1.32-7.28; P < 0.01). In group 90, 1007 alarms (33% false) occurred, whereas in group 85, 395 alarms (28% false) occurred. Introducing a theoretical delay of 15 s in group 90 between crossing the alarm threshold and triggering the alarm would have reduced the number of alarms by 60%. The results of the study suggest that decreasing the alarm limit in an attempt to reduce frequent false alarms may lead to an increase in more relevant episodes of hypoxaemia and setting the LAL at 85% cannot be recommended routinely. Introducing a 15 s delay in group 90 would reduce the number of alarms by the same amount as changing the LAL from 90% to 85%.
在一项前瞻性随机研究中,我们调查了两种任意设定的脉搏血氧饱和度下限报警值(LAL)(90% = 90组,n = 320;85% = 85组,n = 327)对恢复室低氧血症发生率的影响。在90组中,我们计算了通过在报警条件开始与触发报警之间引入时间延迟来消除短暂低脉搏血氧饱和度(SpO2)事件的理论效果。当仅纳入持续超过1分钟的低氧血症发作时,90组中11%的患者出现SpO2≤90%,85组中为20%(相对风险(RR)1.84,置信区间(CI)1.26 - 2.69;P < 0.01)。90组中2%的患者出现SpO2≤85%的低氧血症,85组中为6%(RR 3.10,CI 1.32 - 7.28;P < 0.01)。90组发生了1007次报警(33%为误报),而85组发生了395次报警(28%为误报)。在90组中,在超过报警阈值与触发报警之间引入15秒的理论延迟可使报警次数减少60%。研究结果表明,试图通过降低报警值来减少频繁误报可能会导致更严重的低氧血症发作增加,不建议常规将LAL设定为85%。在90组中引入15秒延迟减少的报警次数与将LAL从90%改为85%相同。