[口腔外科全身麻醉鼻气管插管期间的袖带损伤]

[Cuff damage during naso-tracheal intubation for general anesthesia in oral surgery].

作者信息

Nakamura S, Watanabe T, Hiroi E, Sasaki T, Matsumoto N, Hori T

机构信息

Department of Anesthesiology, Saitama Medical School.

出版信息

Masui. 1997 Nov;46(11):1508-14.

PMID:9404138
Abstract

In our hospital, twenty-one cases of endotracheal tube cuff rupture during naso-tracheal intubation were noted in cases using 725 polyvinyl chloride (PVC) tracheal tubes. We analysed the causes of cuff troubles in these 21 samples of tubes. When the cuffs were inflated, they were not capable of containing the air in most cases. Some cuffs had small holes (described as pinholes), and the others had longer slits on scrape marks and burst. These scrape marks may have been caused by the object with sharp edges such as spina or crista of the nasal septum, or otherwise by the tip of intubation forceps. The cuff material appeared to be slightly hardened in some samples which may be due to the lubrication. We usually lubricated the tube with lidocaine spray or gel formulation and then sometimes placed it in hot water to soften it for avoiding naso-mucosal injury. It is not generally recommended to place tubes in hot water, as this procedure may soften the cuff and make it more suspectible to damage. The clarification is also needed on the use of lidocaine. Although the gel formulation is acceptable, but the spray formulation is known to react with cuff material and make it more susceptible to inducing blistering, pinholes and sudden rupture of PVC cuffs. We conclude that these cuff damages might have occurred from various causes. A main cause must be passing the tube through the narrow nasal turbinate with spina or crista. Other causes could not only be the use of Magill forceps but also lubrication of the tube with lidocaine spray and placing it in hot water.

摘要

在我院,使用725聚氯乙烯(PVC)气管导管经鼻气管插管的病例中,有21例出现气管导管套囊破裂。我们分析了这21根导管套囊出现问题的原因。当套囊充气时,大多数情况下它们无法容纳空气。一些套囊有小孔(称为针孔),另一些在刮痕处有较长的裂缝并破裂。这些刮痕可能是由鼻中隔棘突或嵴等边缘锋利的物体造成的,也可能是由插管钳尖端造成的。在一些样本中,套囊材料似乎稍有硬化,这可能是由于润滑所致。我们通常用利多卡因喷雾或凝胶制剂润滑导管,然后有时将其置于热水中使其软化,以避免鼻黏膜损伤。一般不建议将导管置于热水中,因为此操作可能会使套囊软化,使其更容易受损。利多卡因的使用也需要说明。虽然凝胶制剂是可以接受的,但已知喷雾制剂会与套囊材料发生反应,使其更容易引发PVC套囊起泡、出现针孔和突然破裂。我们得出结论,这些套囊损伤可能是由多种原因造成的。一个主要原因肯定是导管穿过带有棘突或嵴的狭窄鼻鼻甲。其他原因不仅可能是使用麦吉尔钳,还可能是用利多卡因喷雾润滑导管并将其置于热水中。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验