Use of the selective linogram in cardiac tomography quality control.

作者信息

Wallis J W

机构信息

Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, MO 63110, USA.

出版信息

J Nucl Cardiol. 1995 Jul-Aug;2(4):303-8. doi: 10.1016/s1071-3581(05)80074-1.

Abstract

BACKGROUND

Quality control for detection of patient motion is essential in tomographic myocardial imaging. Despite significant limitations, the summation image or conventional "linogram" has long been advocated as a useful image in the detection of vertical motion. In this study a new quality control image entitled the "selective linogram" is proposed to replace the summation image in routine cardiac single-photon emission cardiac tomography (SPECT) quality control. The selective linogram is constructed in a manner somewhat analogous to the sinogram. In the sinogram, each row represents a different projection angle; in the selective linogram each column represents a different projection angle.

METHODS AND RESULTS

After selection of eight motion-free studies from acquisitions at our clinical center, vertical motion of various types (bounces, shifts, and creep) were added to the projection frames. Summation image and selective linogram quality control images from these motion-containing studies and the original motion-free studies were presented in a blinded manner to two observers for scoring of patient motion. The selective linogram was significantly more accurate in allowing detection of vertical motion than was the summation image (accuracy 89% vs 47%).

CONCLUSIONS

The selective linogram image is markedly superior to the summation image for the detection of vertical patient motion during cardiac SPECT. This new technique can be a valuable aid in SPECT quality control.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验