使用Lunar和Hologic仪器对脊柱和股骨近端进行纵向测量的比较。

A comparison of longitudinal measurements in the spine and proximal femur using lunar and hologic instruments.

作者信息

Pocock N A, Noakes K A, Griffiths M, Bhalerao N, Sambrook P N, Eisman J A, Freund J

机构信息

Department of Nuclear Medicine, St. Vincent's Hospital, Sydney, New South Wales, Australia.

出版信息

J Bone Miner Res. 1997 Dec;12(12):2113-8. doi: 10.1359/jbmr.1997.12.12.2113.

Abstract

Absolute values of bone mineral density (BMD), using dual-energy X-ray absorptiometry (DXA), differ between instruments from different manufacturers. Despite these differences, the rates of change calculated from serial measurements on different densitometers have been assumed to be comparable. We compared the change in BMD in 34 subjects in the lumbar spine and at the standard sites in the proximal femur, from measurements performed using a Lunar DPX-L and a Hologic QDR-1000. Measurements were obtained on the same day, and repeated on the same day, on both machines after a mean interval of 4.8 years (range 4.1-6.3 years). There were strong positive correlations between the percentage change in BMD calculated using the two machines in the lumbar spine, trochanteric region, and total proximal femur: r = 0.82, 0.84, and 0.73, respectively (p < 0.0001 at all sites). In the femoral neck and in Ward's triangle, the correlations were not as high: r = 0.55 (p = 0.003) and 0.43 (p = 0.028), respectively. At all sites, despite the significant correlations, the agreement between the two densitometers was not high and there may be significant errors in individual subjects if one uses measurements from one densitometer to predict the change in BMD using the scanner of the other manufacturer. There is less of a problem comparing group data in the lumbar spine and trochanteric region, although errors are still likely to occur in comparing group data of bone loss, calculated using different densitometers, in the other proximal femur sites. In conclusion, the study suggests that caution is necessary in combining bone loss data derived using densitometers from different manufacturers, particularly in the proximal femur. This has important implications for multicenter studies.

摘要

使用双能X线吸收法(DXA)测得的骨密度(BMD)绝对值在不同厂家的仪器之间存在差异。尽管存在这些差异,但不同骨密度仪连续测量得出的变化率仍被认为具有可比性。我们比较了34名受试者腰椎和股骨近端标准部位的骨密度变化,测量分别使用Lunar DPX-L骨密度仪和Hologic QDR-1000骨密度仪进行。在同一天进行测量,并在平均间隔4.8年(范围4.1 - 6.3年)后于同一天在两台仪器上重复测量。两台仪器测得的腰椎、大转子区域和整个股骨近端BMD变化百分比之间存在强正相关:分别为r = 0.82、0.84和0.73(所有部位p < 0.0001)。在股骨颈和沃德三角区,相关性没那么高:分别为r = 0.55(p = 0.003)和0.43(p = 0.028)。在所有部位,尽管相关性显著,但两台骨密度仪之间的一致性并不高,如果使用一台骨密度仪的测量值来预测另一厂家扫描仪的BMD变化,个体受试者可能会有显著误差。在比较腰椎和大转子区域的组数据时问题较小,尽管在比较使用不同骨密度仪计算的股骨近端其他部位骨丢失组数据时仍可能出现误差。总之,该研究表明,在合并使用不同厂家骨密度仪得出的骨丢失数据时需谨慎,尤其是在股骨近端。这对多中心研究具有重要意义。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验