On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems.

作者信息

Borg E, Gröndahl H G

机构信息

Department of Oral Diagnostic Radiology, Faculty of Odontology, Göteborg University, Sweden.

出版信息

Dentomaxillofac Radiol. 1996 Apr;25(2):82-8. doi: 10.1259/dmfr.25.2.9446978.

Abstract

OBJECTIVE

To compare film for intra-oral radiography with two charge-coupled device (CCD) and one storage phosphor system for digital imaging in respect of subjective image quality, detectability of small mass differences and appearance of burn-out effects and blooming phenomena at various exposure times.

METHODS

Dried mandibles with teeth from different areas were radiographed at exposures covering a relative range from 1 to 100. Image quality was subjectively evaluated after image processing, when applicable, using a visual grading scale from 0 to 10. The number of visible holes in an aluminium block was used to measure the detectability of small mass differences. Burn-out effects and blooming were evaluated by measuring widths of roots and of aluminium and plastic cylinders.

RESULTS

Radiographs with the storage phosphor system achieved image quality scores similar to those of film but over a larger exposure range, while CCD images were rated lower and over a smaller range. All holes in the aluminium block were only detected with the storage phosphor system. While the widths of roots were strongly affected by sensor saturation in CCD images and by burn-out in film images, smaller effects were seen with the storage phosphor system. Similar results were obtained with aluminium and plastic cylinders.

CONCLUSIONS

Higher image quality was achieved over a much wider exposure range with the storage phosphor system than with either film or the CCD systems.

摘要

目的

比较口腔内放射摄影胶片与两种电荷耦合器件(CCD)及一种存储磷光体系统在主观图像质量、小质量差异的可检测性以及不同曝光时间下的烧蚀效应和光晕现象方面的数字成像效果。

方法

对来自不同区域的带有牙齿的干燥下颌骨进行放射摄影,曝光范围相对从1至100。在适用时,图像处理后使用0至10的视觉分级量表对图像质量进行主观评估。用铝块中可见孔的数量来测量小质量差异的可检测性。通过测量牙根以及铝和塑料圆柱体的宽度来评估烧蚀效应和光晕。

结果

存储磷光体系统的射线照片在更大的曝光范围内获得了与胶片相似的图像质量评分,而CCD图像的评分较低且范围较小。铝块中的所有孔仅在存储磷光体系统中被检测到。虽然牙根宽度在CCD图像中受传感器饱和以及胶片图像中受烧蚀的强烈影响,但存储磷光体系统的影响较小。铝和塑料圆柱体也得到了类似结果。

结论

与胶片或CCD系统相比,存储磷光体系统在更宽的曝光范围内实现了更高的图像质量。