Suppr超能文献

Evaluation of a new portable system for cord lactate determination.

作者信息

Yam J, Chua S, Razvi K, Arulkumaran S

机构信息

Department of Obstetrics and Gynaecology, National University Hospital, Singapore.

出版信息

Gynecol Obstet Invest. 1998;45(1):29-31. doi: 10.1159/000009919.

Abstract

To evaluate a new reflectance photometer (Accusport) for blood lactate determination, a prospective non-interventional study was performed on blood obtained from the umbilical cord artery immediately after delivery of the baby in 109 consecutive women. For each sample, the results of cord arterial blood lactate values measured with the reflectance photometer were compared with those obtained using a reference method in the laboratory. The values obtained using Accusport were comparable in accuracy with those obtained using a laboratory reference method (Ektachem 9501RC). The instrument provides a simple and reliable method for determining cord arterial blood lactate levels in a small blood sample.