Reproducibility of intraocular pressure and outflow facility measured by pneumatic tonography and Schiotz tonography.

作者信息

Wheeler N C, Lee D A, Cheng Q, Ross W F, Hadjiaghai L

机构信息

Department of Statistics, California Polytechnic State University, San Luis Obispo, USA.

出版信息

J Ocul Pharmacol Ther. 1998 Feb;14(1):5-13. doi: 10.1089/jop.1998.14.5.

Abstract

The purpose of this study was to compare means, variability, and reproducibility in intraocular pressure and in outflow facility measured repeatedly over several days by Bio-Rad pneumatonography and Schiotz tonography in normal human volunteers. For each instrument a measurement of intraocular pressure and one of outflow facility were obtained at each of three repeated examinations of eight healthy subjects in the supine position. For pressure, the Bio-Rad mean was significantly higher by 5.31 mmHg than the Schiotz mean (p < 0.00005). Reproducibility was 71.1% and 59.0% for the Bio-Rad and Schiotz, respectively. For outflow, the Bio-Rad and Schiotz means were not significantly different. Reproducibility was 17.3% and 58.7% for the Bio-Rad and Schiotz, respectively. For measuring pressure, both the Bio-Rad and Schiotz instruments had good reproducibility; for measuring outflow, the Bio-Rad and Schiotz instruments had low and good reproducibility, respectively.