Fink J B, Krause S A, Barrett L, Schaaff D, Alex C G
Department of Veterans Affairs, Edward Hines, Jr. Hospital, Hines, Ill 60141, USA.
Chest. 1998 Feb;113(2):405-11. doi: 10.1378/chest.113.2.405.
To determine the risk of acquiring ventilator-associated pneumonia (VAP) and the impact on costs when extending ventilator circuit change intervals beyond 2 days to 7 and 30 days.
Prospective 4-year review of mechanically ventilated patients.
The respiratory and medical ICUs of an 800-bed tertiary teaching Veterans Affairs hospital.
All adult patients receiving mechanical ventilation from January 1991 through December 1994.
Ventilator circuits with active heated water humidifiers were changed at 2-day intervals during a 2-year control period, followed by 7-day and 30-day intervals (for 1 year each). Heated wire circuits were adopted with the 30-day interval. The rate of VAP per 1,000 ventilator days was calculated for each circuit change interval group. Survival analysis was used to model VAP with ventilator circuit change to determine risk.
During the study period, 637 patients received mechanical ventilation. During the 2 years with 2-day change intervals, the VAP per 1,000 ventilator days was 11.88 (n=343), compared with 3.34 (n=137) and 6.28 (n=157) for 7-day and 30-day change intervals, respectively. The risk of acquiring a VAP for those with a circuit change every 2 days was significantly greater (relative risk, 3.1; p=0.0004; 95% confidence interval, 1.662, 5.812) than those with the 7- and 30-day circuit changes. Extending circuit change intervals reduced supply and labor costs averaging $4,231/yr for each ventilator in use.
Circuit change intervals of 7 and 30 days have lower risks for VAP than the 2-day intervals, yielding substantial reductions in morbidity as well labor and supply costs.
确定将呼吸机回路更换间隔从2天延长至7天和30天对获得呼吸机相关性肺炎(VAP)的风险及成本的影响。
对机械通气患者进行为期4年的前瞻性回顾。
一家拥有800张床位的三级教学退伍军人事务医院的呼吸内科和内科重症监护病房。
1991年1月至1994年12月期间所有接受机械通气的成年患者。
在为期2年的对照期内,每隔2天更换带有主动加热水加湿器的呼吸机回路,随后分别为7天和30天的间隔(各持续1年)。在30天的间隔期采用加热丝回路。计算每个回路更换间隔组每1000个呼吸机日的VAP发生率。采用生存分析对呼吸机回路更换时的VAP进行建模以确定风险。
在研究期间,637例患者接受了机械通气。在每2天更换间隔的2年期间,每1000个呼吸机日的VAP发生率为11.88(n = 343),而7天和30天更换间隔期分别为3.34(n = 137)和6.28(n = 157)。每2天更换一次回路的患者发生VAP的风险显著高于每7天和30天更换回路的患者(相对风险,3.1;p = 0.0004;95%置信区间,1.662, 5.812)。延长回路更换间隔降低了供应和劳动力成本,每台在用呼吸机平均每年节省4231美元。
7天和30天的回路更换间隔比2天的间隔发生VAP的风险更低,可显著降低发病率以及劳动力和供应成本。