Digital cellular telephone interaction with implantable cardioverter-defibrillators.

作者信息

Fetter J G, Ivans V, Benditt D G, Collins J

机构信息

Medtronic, Inc, Minneapolis, Minnesota 55432, USA.

出版信息

J Am Coll Cardiol. 1998 Mar 1;31(3):623-8. doi: 10.1016/s0735-1097(97)00546-9.

Abstract

OBJECTIVES

This study sought to determine, in vivo, whether electromagnetic interference (EMI), generated by North American Digital Communications (NADC)/Time Division Multiple Access-50-Hz (TDMA-50) mobile cellular digital telephone model AT&T 6650, disturbs normal implantable cardioverter-defibrillator (ICD) operation and to verify these observations in vitro by testing a selection of telephones representing worldwide systems.

METHODS

The effects of cellular phone interference on the operation of various models of market-released ICDs from a single manufacturer, Medtronic, Inc., were tested. The in vivo clinical test was undertaken in 41 patients using the AT&T 6650 digital telephone with the NADC/TDMA-50 technology. The in vitro component of the study was examined twofold: 1) antenna generated far field; and 2) analog/digital cellular telephone near field.

RESULTS

None of the ICDs tested in 41 patients were affected by oversensing of the EMI field of the cellular telephones during the in vivo study. Therefore, the binomial upper 95% confidence limit for the failure rate of 0% is 7%. The in vitro antenna-generated field testing showed that telephone modulation frequencies used in the international Global System Mobile and TDMA-50 cellular telephone technologies did not result in ICD sensing interference at the predicted electric field intensity. The in vitro near field tests were performed using both analog and digital cellular telephones in service, or in the test mode, and indicated no interaction with normal operation. However, the static magnetic field generated by the cellular telephone placed over the ICD at a distance < or = 0.5 cm will activate the internal reed switch, resulting in temporary suspension of ventricular tachycardia and fibrillation detection.

CONCLUSIONS

We conclude that TDMA-50 cellular telephones did not interfere with these types of ICDs. However, we recommend that the patient not carry or place the digital cellular telephone within 15 cm (6 in.) of the ICD.