Coselli J S, LeMaire S A, Ledesma D F, Ohtsubo S, Tayama E, Nosé Y
Department of Surgery, Baylor College of Medicine, Houston, Tex. 77030, USA.
J Vasc Surg. 1998 Feb;27(2):378-83. doi: 10.1016/s0741-5214(98)70372-6.
Several centers use atriodistal bypass (ADB) as a protective adjunct against distal ischemia during extensive thoracoabdominal aortic aneurysm (TAAA) repair. Most current ADB circuits use indirect-drive centrifugal pumps. The purpose of this report is to describe our initial clinical experience with the Nikkiso pump, a more compact direct-drive centrifugal pump recently developed at Baylor, for ADB during TAAA repair.
The Nikkiso pump was used for ADB perfusion in 10 consecutive patients during graft repair of TAAAs (six Crawford extent I and four extent II). Two patients had aortic dissection. In the four patients who had extent II repairs, selective renal and visceral perfusion was also performed with the Nikkiso pump.
No mechanical pump malfunctions or adverse events related to the device occurred. All 10 patients survived and were discharged from the hospital. No patient had paraplegia after surgery. Two patients had delayed lower extremity weakness after undergoing extent I repairs; both recovered and were ambulating at the time of discharge. No complications were associated with bleeding or cerebral, respiratory, renal, or hepatic function.
Our initial experience with the Nikkiso centrifugal pump during TAAA repair demonstrated excellent pump function that provided sufficient flow for both distal aortic and selective organ perfusion. The prevention of permanent spinal cord injury and distal organ failure was successful in this group.
多个中心在广泛胸腹主动脉瘤(TAAA)修复术中使用心房远端旁路(ADB)作为预防远端缺血的辅助措施。目前大多数ADB回路使用间接驱动离心泵。本报告的目的是描述我们使用Nikkiso泵的初步临床经验,这是一种最近在贝勒大学开发的更紧凑的直接驱动离心泵,用于TAAA修复术中的ADB。
在连续10例TAAA移植修复患者(6例克劳福德I型和4例II型)中使用Nikkiso泵进行ADB灌注。2例患者患有主动脉夹层。在4例进行II型修复的患者中,也使用Nikkiso泵进行了选择性肾和内脏灌注。
未发生与该设备相关的机械泵故障或不良事件。所有10例患者均存活并出院。术后无患者发生截瘫。2例I型修复患者术后出现下肢延迟性无力;两人均康复,出院时可行走。未发生与出血或脑、呼吸、肾或肝功能相关的并发症。
我们在TAAA修复术中使用Nikkiso离心泵的初步经验表明,该泵功能出色,可为远端主动脉和选择性器官灌注提供足够的流量。该组成功预防了永久性脊髓损伤和远端器官衰竭。