健康受试者的压力测痛法:检查者间的变异性。

Pressure algometry in healthy subjects: inter-examiner variability.

作者信息

Antonaci F, Sand T, Lucas G A

机构信息

University Centre for Adaptive Disorders and Headache (UCADH), Section of Pavia I, C. Mondino Foundation, Pavia, Italy.

出版信息

Scand J Rehabil Med. 1998 Mar;30(1):3-8. doi: 10.1080/003655098444255.

Abstract

The purpose of this study was to estimate inter-examiner reliability of head and neck algometry. Pain perception thresholds were assessed with a mechanical pressure algometer in 21 healthy individuals. Thresholds were assessed at 13 symmetrical points on each side of the head and neck, at the deltoid muscle and at the median finger. The pressure range of the instrument proved insufficient to study the pain perception threshold on the finger, however. Two different examiners carried out one or two examinations in each subject during one day. The sequence of investigations was varied randomly. The inter-examiner reliability was found to be good, with a mean intra-class correlation coefficient (ICC) of 0.75. Intra-examiner reproducibility was excellent (mean ICC = 0.84). The mean inter-examiner coefficient of variation was 18.7%, while the mean coefficient of repeatability (CR) was 1.60 kg/cm2. In comparison, the mean intra-examiner coefficient of variation was 15% while the mean CR was 1.29 kg/cm2. Statistically significant differences between examiners were found for the frontal point (p < 0.01), while a trend towards lower thresholds in one of the two observers was seen in 10 of the 13 non-significant points. Inter-examiner reliability of side differences was excellent, with CR = 1.23 kg/cm2. In conclusion, manual algometry with a rather inexpensive mechanical device has a good to excellent inter-rater reliability. When studying patients, however, the possible bias introduced by different examiners should be taken into account, both regarding study design and data analysis.

摘要

本研究的目的是评估头颈压力测定法中检查者间的可靠性。使用机械压力测定仪对21名健康个体的疼痛感知阈值进行了评估。在头颈部两侧的13个对称点、三角肌和中指处评估阈值。然而,该仪器的压力范围被证明不足以研究手指上的疼痛感知阈值。两名不同的检查者在一天内对每个受试者进行了一到两次检查。检查顺序随机变化。发现检查者间的可靠性良好,组内相关系数(ICC)的平均值为0.75。检查者内的可重复性极佳(平均ICC = 0.84)。检查者间变异系数的平均值为18.7%,而重复性系数(CR)的平均值为1.60 kg/cm²。相比之下,检查者内变异系数的平均值为15%,而平均CR为1.29 kg/cm²。在额点发现检查者之间存在统计学显著差异(p < 0.01),而在13个无显著差异的点中的10个点上,两名观察者之一出现阈值较低的趋势。两侧差异的检查者间可靠性极佳,CR = 1.23 kg/cm²。总之,使用相当便宜的机械设备进行手动压力测定法具有良好到极佳的评分者间可靠性。然而,在研究患者时,在研究设计和数据分析方面,都应考虑不同检查者可能引入的偏差。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验