Suppression of an external demand pacemaker by diaphragmatic myopotentials: a sign of electrode perforation?

作者信息

El Gamal M, Van Gelder B

出版信息

Pacing Clin Electrophysiol. 1979 Mar;2(2):191-5. doi: 10.1111/j.1540-8159.1979.tb05200.x.

Abstract

Transient pacemaker inhibition was observed in a patient with an external bipolar demand pulse generator. Recordings of the proximal, distal and bipolar electrograms showed extracardiac myopotentials only in the unipolar distal and bipolar configuration. These myopotentials which caused inhibition of the pulse generator were related to respiration and other maneuvers that resulted in active contraction of the diaphragm, thus proving their diaphragmatic origin.