Demonstration of a possible source of error with an electric pulp tester.

作者信息

Myers J W

机构信息

Indiana University School of Dentistry, USA.

出版信息

J Endod. 1998 Mar;24(3):199-200. doi: 10.1016/S0099-2399(98)80184-2.

Abstract

The purpose of this study was to determine whether electrical current can travel between adjacent teeth through contacting interproximal amalgam restorations. Twenty-two extracted human premolars were restored with class II amalgam restorations. They were then mounted in pairs in self-curing resin, simulating adjacent teeth. Only their restorations were allowed to contact. An electrical multimeter was used to supply electrical current and to measure any current passing through the mounted teeth. One current measurement was made from the buccal enamel of a tooth to the occlusal surface of the filling in that tooth. A second measurement was made from the buccal enamel of the same tooth to the occlusal surface of the adjacent filling. These two measurements were made for all 11 trials. Statistically, all current that entered the first tooth passed to the adjacent tooth. This finding suggests a potential source of error when using an electric pulp tester.

摘要

本研究的目的是确定电流是否能通过相邻牙齿间相互接触的邻面汞合金修复体在相邻牙齿之间传导。二十二颗拔除的人类前磨牙用II类汞合金修复体进行修复。然后将它们成对安装在自凝树脂中,模拟相邻牙齿。仅允许它们的修复体相互接触。使用电子万用表来供应电流并测量通过安装好的牙齿的任何电流。一次电流测量是从一颗牙齿的颊侧牙釉质到该牙齿充填物的咬合面。第二次测量是从同一颗牙齿的颊侧牙釉质到相邻充填物的咬合面。对所有11次试验都进行了这两种测量。从统计学角度来看,所有进入第一颗牙齿的电流都传导到了相邻牙齿。这一发现表明在使用牙髓电活力测试仪时存在潜在的误差来源。