Frey B, Butt W
Intensive Care Unit, Royal Children's Hospital, Melbourne, Victoria, Australia.
Intensive Care Med. 1998 Mar;24(3):242-6. doi: 10.1007/s001340050557.
To evaluate the relationship between intraarterial measured pulsus paradoxus (PP) and photoplethysmographic wave changes.
Tertiary pediatric intensive care unit.
62 nonintubated children with or without respiratory disorders.
Prospective, clinical study.
Simultaneous paper recordings of photoplethysmographic wave, arterial blood pressure, breathing cycle and electrocardiogram. The respiratory dependent changes of the plethysmographic respiratory wave (delta pleth, mm) were defined as the difference between the highest value of the upper peak of the wave (in expiration) and the lowest value of the upper peak (in inspiration). In each patient, ten consecutive breaths were averaged for analysis. Five recordings could not be evaluated (movement artifacts). In 57 children (median age 2.4 years, range 7 days to 17 years), the photoplethysmographic fluctuations (delta pleth, mm) correlated with PP (mm Hg). r = 0.85; 95% confidence interval (CI), 0.76 to 0.91. The sensitivity to detect a PP of > 10 mmHg with a plethysmographic fluctuation of > 8 mm was 89% (95% CI, 77 to 100%) and the specificity was 90% (95% CI, 79 to 100%).
Pulse oximetry appears to be a rapid and easily performed, noninvasive method for the objective estimation of the degree of PP.
评估动脉内测压的奇脉(PP)与光电容积脉搏波变化之间的关系。
三级儿科重症监护病房。
62例有或无呼吸系统疾病的非插管儿童。
前瞻性临床研究。
同步记录光电容积脉搏波、动脉血压、呼吸周期和心电图。光电容积脉搏波的呼吸依赖性变化(δ容积,mm)定义为波的上峰最高值(呼气时)与上峰最低值(吸气时)之间的差值。对每位患者连续10次呼吸进行平均分析。5份记录因运动伪影无法评估。在57名儿童(中位年龄2.4岁,范围7天至17岁)中,光电容积脉搏波波动(δ容积,mm)与PP(mmHg)相关。r = 0.85;95%置信区间(CI)为0.76至0.91。光电容积脉搏波波动>8 mm时检测PP>10 mmHg的敏感性为89%(95%CI,77至100%),特异性为90%(95%CI,79至100%)。
脉搏血氧饱和度测定似乎是一种快速、易于实施的无创方法,可用于客观评估奇脉程度。