Lagergren H, Edhag O, Wahlberg I
J Thorac Cardiovasc Surg. 1976 Aug;72(2):259-64.
Threshold curves with large and small surface intracardiac pacemaker electrodes are compared. The 2 msec. impulse threshold with a 47 sq. mm. electrode was 3.6 v. (4.3 mA.) on the fourteenth postoperative day, when it reached its maximum, and 2.8 v. (3.1 mA.) one month after the operation. These values were 45 and 30 per cent lower with a 6 sq. mm. electrode. Thresholds increased by about 20 per cent when the impulse duration was shortened from 2 to 0.5 msec. The small surface electrode consumed about 35 per cent less current than the 47 sq. mm. one. A newly designed large area-small surface electrode with the shape of an open cage, seems to have the advantages of less increase in postoperative thresholds and good attachment to the endocardial wall.
比较了大小不同的心内表面起搏器电极的阈值曲线。术后第14天,当47平方毫米电极的2毫秒脉冲阈值达到最大值时为3.6伏(4.3毫安),术后1个月为2.8伏(3.1毫安)。对于6平方毫米的电极,这些值分别低45%和30%。当脉冲持续时间从2毫秒缩短到0.5毫秒时,阈值增加约20%。小表面电极消耗的电流比47平方毫米的电极少约35%。一种新设计的呈开放式笼状的大面积-小表面电极,似乎具有术后阈值增加较少以及与心内膜壁附着良好的优点。