Jiang S B, Ayyangar K M
Department of Radiation Therapy, Medical College of Ohio, Toledo 43614-2598, USA.
Med Phys. 1998 May;25(5):668-75. doi: 10.1118/1.598250.
Recently the compensator has been shown to be an in expensive and reliable dose delivery device for photon beam intensity-modulated radiation therapy (IMRT). The goal of IMRT compensator design is to produce an optimized primary fluence profile at the patient's surface obtained from the optimization procedure. In this paper some of the problems associated with IMRT compensator design, specifically the beam perturbations caused by the compensator, are discussed. A simple formula is derived to calculate the optimal compensator thickness profile from an optimized primary fluence profile. The change of characteristics of a 6 MV beam caused by the introduction of cerrobend compensators in the beam is investigated using OMEGA Monte Carlo codes. It is found that the compensator significantly changes the energy spectrum and the mean energy of the primary photons at the patient's surface. However, beam hardening does not have as significant an effect on the percent depth dose as it does on the energy spectrum. We conclude that in most situations the beam hardening effect can be ignored during compensator design and dose calculation. The influence of the compensator on the contaminant electron buildup dose is found to be small and independent of the compensator thickness of interest. Therefore, it can be ignored in the compensator design and included as a correction into the final dose distribution. The scattered photons from the compensator are found to have no effect on the surface dose. These photons produce a uniform low fluence distribution at the patient's surface, which is independent of compensator shape. This is also true for very large fields and extremely asymmetric and nonuniform compensator thickness profiles. Compared to the primary photons, the scattered photons have much larger angular spread and similar energy spectrum at the patient's surface. These characteristics allow the compensator thickness profile and the dose distribution to be calculated from the optimized fluence profile of primary photons, without considering the scattered photons.
最近,补偿器已被证明是一种用于光子束调强放射治疗(IMRT)的廉价且可靠的剂量输送装置。IMRT补偿器设计的目标是在优化程序获得的患者表面产生优化的初始注量分布。本文讨论了与IMRT补偿器设计相关的一些问题,特别是补偿器引起的射束扰动。推导出一个简单公式,用于从优化的初始注量分布计算最佳补偿器厚度分布。使用OMEGA蒙特卡罗代码研究了在射束中引入铈基合金补偿器导致的6兆伏射束特性变化。发现补偿器显著改变了患者表面初始光子的能谱和平均能量。然而,射束硬化对百分深度剂量的影响不如对能谱的影响那么显著。我们得出结论,在大多数情况下,在补偿器设计和剂量计算过程中可以忽略射束硬化效应。发现补偿器对污染电子积累剂量的影响很小,且与所关注的补偿器厚度无关。因此,在补偿器设计中可以忽略它,并作为校正纳入最终剂量分布。发现来自补偿器的散射光子对表面剂量没有影响。这些光子在患者表面产生均匀的低注量分布,这与补偿器形状无关。对于非常大的射野以及极其不对称和不均匀的补偿器厚度分布也是如此。与初始光子相比,散射光子在患者表面具有大得多的角展宽和相似的能谱。这些特性使得可以根据初始光子的优化注量分布计算补偿器厚度分布和剂量分布,而无需考虑散射光子。