Lin H, Xu H, Zhang X, de Groot K
Dental Materials Lab., Stomatological College of Beijing Medical University, People's Republic of China.
J Biomed Mater Res. 1998 Summer;43(2):113-22. doi: 10.1002/(sici)1097-4636(199822)43:2<113::aid-jbm5>3.0.co;2-r.
Tensile strength testing of the interface between bone and a plasma-sprayed hydroxyapatite (HA) coating-Ti-6Al-4V (implant I) and an HA coating-Ti (implant II) was performed. The bone-implant interface and tensile failure mode were evaluated by light microscopy (LM), SEM-energy dispersive X-ray analysis, and backscattered electron imaging. The crystallinity of the HA coating of implant I and implant II was 62 and 40%, respectively. Eleven semidisk implants of each type were inserted in the left tibia of 22 rabbits in each period. After 2, 4, 8, and 16 weeks, the tensile strength of the bone-implant I was 0.86, 1.32, 1.10, and 0.92 MPa, respectively; the value of the bone-implant II was 0.66, 0.92, 0.84, and 1.12 MPa, respectively. No significant difference was found in tensile strength between implant I and implant II, and between the data of different periods. LM and SEM revealed that the two types of coatings behave the same in bone. Failure after the tensile test mainly occurred at the bony tissue at 2 and 4 weeks, inside the HA coating layer at 8 weeks, and at the HA coating-metal substrate interface at 16 weeks. Degradation and delamination of the HA coating continued with time, resulting in the interfacial tensile strength not increasing with time. Thus, we concluded that differences between the two implants, mainly the difference in crystallinity of the HA coating (62 vs. 40%), have no significant influence on bone bonding ability, bone bonding strength, tensile failure mode, and degradation of the coating.
对骨与等离子喷涂羟基磷灰石(HA)涂层 - Ti - 6Al - 4V(植入物I)以及HA涂层 - Ti(植入物II)之间的界面进行了拉伸强度测试。通过光学显微镜(LM)、扫描电子显微镜 - 能量色散X射线分析和背散射电子成像对骨 - 植入物界面和拉伸破坏模式进行了评估。植入物I和植入物II的HA涂层的结晶度分别为62%和40%。在每个时间段,将每种类型的11个半圆形植入物插入22只兔子的左胫骨中。在2、4、8和16周后,骨 - 植入物I的拉伸强度分别为0.86、1.32、1.10和0.92MPa;骨 - 植入物II的值分别为0.66、0.92、0.84和1.12MPa。在植入物I和植入物II之间以及不同时间段的数据之间,拉伸强度没有发现显著差异。光学显微镜和扫描电子显微镜显示,两种涂层在骨中的表现相同。拉伸试验后的破坏主要发生在2周和4周时的骨组织处,8周时在HA涂层内部,16周时在HA涂层 - 金属基底界面处。HA涂层的降解和分层随时间持续,导致界面拉伸强度没有随时间增加。因此,我们得出结论,两种植入物之间的差异,主要是HA涂层结晶度的差异(62%对40%),对骨结合能力、骨结合强度、拉伸破坏模式和涂层降解没有显著影响。