First-pass radionuclide angiography using iodine-123 myocardial tracers and a multicrystal gamma camera.

作者信息

Tadamura E, Kitano H, Kudoh T, Hattori N, Inubushi M, Nishimura K, Masuda I, Inada H, Hosokawa R, Nohara R, Konishi J

机构信息

Department of Nuclear Medicine, Kyoto University Faculty of Medicine, Japan.

出版信息

J Nucl Med. 1998 Jun;39(6):938-44.

PMID:9627322
Abstract

UNLABELLED

The purpose of this study was to validate the accuracy of the assessment of ventricular function by first-pass radionuclide angiography (FPRNA) with 123I myocardial tracers and a multicrystal gamma camera.

METHODS

Left ventricular ejection fraction (LVEF) and right ventricular ejection fraction were measured in 69 patients by FPRNA using 123I myocardial tracers (126 +/- 7 MBq) and 99mTc tracers (541 +/- 141 MBq) on a multicrystal gamma camera with a high-sensitivity collimator. For 44 patients, ejection fraction values measured by 123I-FPRNA were compared to those estimated by equilibrium radionuclide angiography (ERNA). Visual wall-motion analysis was also performed to judge clinical acceptability of 123I-FPRNA images for identification of wall-motion abnormality.

RESULTS

Mean LVEFs (%) estimated by 123I-FPRNA and by 99mTc-FPRNA were 49.6 +/- 13.6 and 49.1 +/- 14.1, respectively (nonsignificant p value). An excellent correlation was found between LVEFs estimated by 123I-FPRNA and 99mTc-FPRNA (r = 0.96, s.e.e. = 1.9%). Values of LVEF measured by 123I-FPRNA also demonstrated excellent correlation with those measured by ERNA (r = 0.95, s.e.e. = 2.2%). A good correlation was also noted between right ventricular ejection fractions measured by 123I-FPRNA and 99mTc-FPRNA (r = 0.72, s.e.e. = 4.0%). The Spearman rank correlation coefficient between 123I-FPRNA and ERNA wall-motion scores was 0.87 (n = 135, p < 0.001).

CONCLUSION

Resting ventricular function can be reliably measured with 123I-FPRNA in combination with a multicrystal gamma camera. This indicates that the assessment of ventricular function is feasible in conjunction with 123I myocardial imaging without an increase in cost or radiation dose to patients.

摘要

未标注

本研究的目的是验证使用123I心肌示踪剂和多晶体γ相机进行首次通过放射性核素血管造影(FPRNA)评估心室功能的准确性。

方法

在一台配备高灵敏度准直器的多晶体γ相机上,使用123I心肌示踪剂(126±7MBq)和99mTc示踪剂(541±141MBq),通过FPRNA对69例患者的左心室射血分数(LVEF)和右心室射血分数进行测量。对44例患者,将123I-FPRNA测量的射血分数值与平衡放射性核素血管造影(ERNA)估计的值进行比较。还进行了视觉壁运动分析,以判断123I-FPRNA图像用于识别壁运动异常的临床可接受性。

结果

123I-FPRNA和99mTc-FPRNA估计的平均LVEF(%)分别为49.6±13.6和49.1±14.1(p值无显著性差异)。123I-FPRNA和99mTc-FPRNA估计的LVEF之间发现了极好的相关性(r = 0.96,标准误差 = 1.9%)。123I-FPRNA测量的LVEF值与ERNA测量的值也显示出极好的相关性(r = 0.95,标准误差 = 2.2%)。123I-FPRNA和99mTc-FPRNA测量的右心室射血分数之间也发现了良好的相关性(r = 0.72,标准误差 = 4.0%)。123I-FPRNA和ERNA壁运动评分之间的Spearman等级相关系数为0.87(n = 135,p < 0.001)。

结论

123I-FPRNA结合多晶体γ相机可可靠地测量静息心室功能。这表明在不增加患者成本或辐射剂量的情况下,结合123I心肌成像评估心室功能是可行的。