Can an oxygenator design potentially contribute to air embolism in cardiopulmonary bypass? A novel method for the determination of the air removal capabilities of neonatal membrane oxygenators.

作者信息

De Somer F, Dierickx P, Dujardin D, Verdonck P, Van Nooten G

机构信息

Institute of Biomedical Technology, University of Gent, Belgium.

出版信息

Perfusion. 1998 May;13(3):157-63. doi: 10.1177/026765919801300302.

Abstract

At present, air handling of a membrane oxygenator is generally studied by using an ultrasonic sound bubble counter. However, this is not a quantitative method and it does not give any information on where air was entrapped in the oxygenator and if it eventually was removed through the membrane for gas exchange. The study presented here gives a novel technique for the determination of the air-handling characteristics of a membrane oxygenator. The study aimed at defining not only the amount of air released by the oxygenator, but also the amount of air trapped within the oxygenator and/or removed through the gas exchange membrane. Two neonatal membrane oxygenators without the use of an arterial filter were investigated: the Polystan Microsafe and the Dideco Lilliput. Although the air trap function of both oxygenators when challenged with a bolus of air was similar, the Microsafe obtained this effect mainly by capturing the air in the heat exchanger compartment while the Lilliput did remove a large amount of air through the membrane. In conclusion, the difference in trap function was most striking during continuous infusion of air. Immediate contact with a microporous membrane, avoidance of high velocities within the oxygenator, pressure drop, transit time and construction of the fibre mat all contribute to the air-handling characteristics of a membrane oxygenator.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验