Effectiveness of the Gaussian beam model in predicting SAR distributions from the lucite cone applicator.

作者信息

Rietveld P J, Lumori M L, Hand J W, Prior M V, Van der Zee J, Van Rhoon G C

机构信息

University Hospital Rotterdam/Daniel den Hoed Cancer Center, Department of Radiation Oncology, The Netherlands.

出版信息

Int J Hyperthermia. 1998 May-Jun;14(3):293-308. doi: 10.3109/02656739809018234.

Abstract

The Gaussian beam model (GBM) has been shown to be a successful tool in the development of the current sheet applicator. As a result, the effectiveness of the GBM is investigated in single and dual array applications of the lucite cone applicator (LCA). The LCA is a modified water-filled waveguide applicator with an improved effective field size (EFS > 64 cm2, aperture 10 cm x 10 cm). The GB-source parameters were calculated from the emanating E-field of a single LCA. The SAR distribution from a single LCA was measured by E-field scanning and thermographic (TG) imaging, and compared with the GB-predicted SAR distribution. Deviations in the principal planes were found to be less than 5%. TG-measured and GB-predicted SAR distributions from three different dual LCA configurations were compared and evaluated. When water was used as intermedium between LCAs and phantom, a maximum SAR difference of 27% was calculated. In the absence of water as intermedium, this difference increased to 44%. These large deviations were only found in areas where the measured SAR distribution was disturbed due to antenna interactions. The average SAR differences with and without water as intermedium were 7% respectively 11%, indicating that the GBM can provide good qualitative information about the SAR distribution of dual LCA-arrays.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验