van der Vis H M, Zwartelé R, Schüller H M, Doets H K, Marti R K
Department of Orthopaedic Surgery, Academic Medical Centre, University of Amsterdam, The Netherlands.
Acta Orthop Scand. 1998 Jun;69(3):248-52. doi: 10.3109/17453679809000924.
We measured radiographically the polyethylene socket wear in 34 hip prostheses with a 32 mm Al2O3-ceramic head with a rotating bearing and in 37 prostheses with a "fixed" ceramic head. The mean follow-up was 12 years in both groups. The mean annual linear wear of the polyethylene was 0.034 and 0.069 mm, respectively, (Mann-Whitney U-test p < 0.0001) in the "rotation" and the "fixed" group. A rotating bearing between the head and neck in a modular total hip system seems to reduce socket wear as compared to fixed taper junctions.
我们通过影像学测量了34个采用32毫米氧化铝陶瓷股骨头旋转轴承的髋关节假体和37个采用“固定”陶瓷股骨头的髋关节假体中聚乙烯髋臼杯的磨损情况。两组的平均随访时间均为12年。“旋转”组和“固定”组中聚乙烯的年平均线性磨损分别为0.034毫米和0.069毫米(曼-惠特尼U检验,p < 0.0001)。与固定锥形连接相比,模块化全髋关节系统中股骨头与股骨颈之间的旋转轴承似乎能减少髋臼杯磨损。