Accuracy and precision of blood pressure determination with the Finapres: an overview using re-sampling statistics.

作者信息

Silke B, McAuley D

机构信息

Therapeutics and Pharmacology, The Whitla Division of Medicine, The Queen's University of Belfast, Northern Ireland.

出版信息

J Hum Hypertens. 1998 Jun;12(6):403-9. doi: 10.1038/sj.jhh.1000600.

Abstract

The Finapres non-invasive blood pressure (BP) monitor uses the method of Penaz to indirectly record the arterial waveform; studies on its accuracy have suggested little systematic bias vs intra-arterial pressure (IAP) but substantial variability. Inconsistency between studies, in respect of the magnitude, direction and variance of bias, was described in the validation studies against the direct IAP. We have employed a novel re-sampling statistical method to combine the data from 20 published studies; a robust overall estimate of the accuracy and precision of the Finapres was thereby obtained. Based on 449 patients and 4490 re-samples, the average Finapres systolic bias (IAP- Finapres) was 2.2 mm Hg (s.d.+/-12.4) with limits of agreement (bias+/-2 s.d.) of -22.6 and 26.9 mm Hg. The average precision was 12.1 mm Hg (s.d.+/-8.4). The Finapres diastolic bias was -0.3 mm Hg (s.d.+/-7.9) with the limits of agreement of -16.1 and 15.5 mm Hg. The average precision was 7.6 mm Hg (s.d.+/-5.3). The average Finapres mean arterial pressure bias was 2.1 mm Hg (s.d.+/-8.6) with precision of 7.6 mm Hg (s.d.+/-5.3). The calculated percentage of Finapres systolic values expected to fall within +/-5 or +/-10 of the direct intra-arterial pressure was 35.9% and 73.1%, respectively. The calculated precision of the Finapres systolic pressure between 0-5 mm Hg was 1.6% and between 0-10 mm Hg 36.4%. The comparable values for Finapres diastolic BP for accuracy were 63.5% and 92.8% and for precision 23.1% and 79.2%. The Finapres device can provide an accurate estimate of diastolic and mean arterial pressure compared with the intra-arterial record; the apparent inaccuracy of the Finapres systolic pressure may have a physiological explanation. When the Finapres device is used in experimental or in clinical situations, then calibration against a reliable reference arterial pressure is desirable to obviate the possibility of an 'offset' error.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验