[Recording of ventricular pressure by conventional catheter manometer systems. Efficiency of several combinations of conventional catheters, modern transducers and catheter-flush systems (author's transl)].

作者信息

Hellige G

出版信息

Basic Res Cardiol. 1976 Jul-Aug;71(4):389-406. doi: 10.1007/BF01910777.

Abstract

The experimentally in vitro determined dynamic response characteristics of 38 catheter manometer systems were uniform in the worst case to 5 c.p.s. and optimally to 26 c.p.s. Accordingly, some systems are only satisfactory for ordinary pressure recording in cardiac rest, while better systems record dp/dt correct up to moderate inotropic stimulation of the heart. In the frequency range of uniform response (amplitude error less +/- 5%) the phase distortion is also negligible. In clinical application the investigator is often restricted to special type of cardiac catheter. In this case a low compliant transducer yields superior results. In all examined systems the combination with MSD 10 transducers is best, whereas the combination with P 23 Db transducers leads to minimal results. An inadequate system for recording ventricular pressure pulses leads in most cases to overestimations of dp/dtmax. The use of low frequency pass filters to attenuate higher frequency artefacts is, under clinical conditions, not suitable for extending the range of uniform frequency response. The dynamic response of 14 catheter manometer systems with two types of continuous self flush units was determined. The use of the P 37 flush unit in combination with small internal diameter catheters leads to serious error in ordinary pressure recording, due to amplitude distortion of the lower harmonics. The frequency response characteristics of the combination of an Intraflow flush system and MSD 10 transducer was similar to the non-flushing P 23 Db transducer feature.

摘要

相似文献

4
Time-dependent pressure distortion in a catheter-transducer system: correction by fast flush.
Anesthesiology. 2000 Jan;92(1):208-18. doi: 10.1097/00000542-200001000-00033.
6
The fast flush test: evaluation in radial artery catheter-manometer systems.
Clin Phys Physiol Meas. 1991 Aug;12(3):247-52. doi: 10.1088/0143-0815/12/3/005.
10
Experimental analysis of catheter-manometer systems in vitro and in vivo.
Anesthesiology. 1989 Dec;71(6):893-906. doi: 10.1097/00000542-198912000-00013.

本文引用的文献

3
Catheter-flush system for continuous monitoring of central arterial pulse waveform.
J Appl Physiol. 1970 Dec;29(6):911-3. doi: 10.1152/jappl.1970.29.6.911.
4
The contractile state of the hypertrophied left ventricular myocardium in aortic stenosis.
Am Heart J. 1970 May;79(5):587-602. doi: 10.1016/0002-8703(70)90278-4.
7
V max as an index of contractile state in man.
Circulation. 1971 Sep;44(3):486-7. doi: 10.1161/01.cir.44.3.486.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验