Frequency response evaluation of radial artery catheter-manometer systems: sinusoidal frequency analysis versus flush method.

作者信息

Schwid H A

机构信息

Department of Anesthesiology, University of Washington, Seattle.

出版信息

J Clin Monit. 1988 Jul;4(3):181-5. doi: 10.1007/BF01621814.

Abstract

It is well recognized that catheter-manometer systems significantly distort direct radial artery pressure measurements. Sinusoidal frequency analysis and the flush method of assessing the degree of distortion caused by the monitoring system were compared to determine whether these two methods agree in the estimation of natural frequency and damping coefficient. The frequency response of 30 radial artery catheter-manometer systems used for intensive-care unit patients was measured by the flush method and sinusoidal frequency analysis. The monitoring system consisted of a 20-gauge cannula, 150-cm pressure tubing, two plastic stopcocks, a continuous infusion device with fast flush valve, an American Edwards dome, a Hewlett-Packard quartz transducer, and a Hewlett-Packard blood pressure amplifier. Sinusoidal frequency analysis demonstrated second-order underdamped response for all 30 catheter-manometer systems. No secondary resonance peaks were observed up to a frequency of 200 Hz. The measured frequency response demonstrated that the average catheter-manometer system in use in our intensive care unit would cause significant distortion of the radial artery pressure, with the mean natural frequency (fn) of 14.7 +/- 3.7 Hz and the mean damping coefficient (zeta) of 0.24 +/- 0.07. Although the 30 monitoring systems had identical configurations and visible bubbles were carefully removed, a wide range of frequency responses was found (fn = 10.2 to 25.3; zeta = 0.15 to 0.44).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验