Suppr超能文献

气腹过滤器在腹腔镜手术中的效用。

Utility of insufflation filters in laparoscopic surgery.

作者信息

Frankel S M, Fitzgerald R L, Sackier J M

机构信息

Department of Surgery, 2150 Pennsylvania Avenue, NW, Suite 6-B, The George Washington University Medical Center, Washington DC 20037, USA.

出版信息

Surg Endosc. 1998 Sep;12(9):1137-40. doi: 10.1007/s004649900800.

Abstract

OBJECTIVES

To determine whether filters, regularly used as part of the insufflator tubing during laparoscopic surgery, trap microbial and particulate matter from CO2 tanks, thus preventing passage from one patient to another.

METHODS

A total of 67 used filters were collected from 17 CO2 tanks and six insufflation machines at three local hospitals, and sterile unused filters were used as controls. The used filters were distributed equally and sequentially into three groups: Group I-viewed under a dissecting microscope for particulate matter; group II-examined by mass spectrometry for contamination with oils and other impurities; group III-incubated on sheep blood agar plates and evaluated for growth of microorganisms.

RESULTS

Negative. Used filters were indistinguishable by all parameters from controls.

CONCLUSIONS

This limited study suggests filters now used in laparoscopic surgery fail to trap microbes or particulate matter. The question remains whether tank waste is absent or these filters fail to trap waste matter.

摘要

目的

确定在腹腔镜手术期间作为气腹机管道一部分经常使用的过滤器是否能捕获来自二氧化碳罐的微生物和颗粒物,从而防止其在患者之间传播。

方法

从当地三家医院的17个二氧化碳罐和六台气腹机中收集了总共67个用过的过滤器,并将未使用的无菌过滤器用作对照。将用过的过滤器平均且依次分为三组:第一组——在解剖显微镜下观察颗粒物;第二组——通过质谱法检测是否被油类和其他杂质污染;第三组——在绵羊血琼脂平板上培养并评估微生物生长情况。

结果

阴性。用过的过滤器在所有参数上与对照无差异。

结论

这项有限的研究表明,目前腹腔镜手术中使用的过滤器无法捕获微生物或颗粒物。问题仍然是罐内是否没有废物,还是这些过滤器无法捕获废物。