Suppr超能文献

近期电子剂量测定实用规范的实验评估。

An experimental evaluation of recent electron dosimetry codes of practice.

作者信息

Nisbet A, Thwaites D I, Nahum A E, Pitchford W G

机构信息

Department of Medical Physics, Leicester Royal Infirmary NHS Trust, UK.

出版信息

Phys Med Biol. 1998 Aug;43(8):1999-2014. doi: 10.1088/0031-9155/43/8/001.

Abstract

As from the 1 January 1997, the recent IPEMB code of practice for electron dosimetry is the recommended protocol for electron beam dosimetry in the UK, replacing the previous HPA code of practice and its IPSM addendum. New recommendations for electron beam dosimetry have also been formulated recently by the AAPM and the IAEA on the use of parallel-plate ionization chambers in high-energy electron beams. Against this background, the procedures recommended in each of these codes of practice have been followed from intercomparison of the field instrument ionization chamber with a secondary standard through to the determination of absorbed dose at the reference position in the electron beam. Absorbed doses have been determined for a number of electron beam energies ranging from nominal 5 MeV through to 17 MeV, and for four different types of field instrument ionization chamber: an NE2571 graphite walled cylindrical chamber; an NACP parallel-plate chamber; a Markus parallel-plate chamber; and a Roos parallel-plate chamber. The differences in the determination of absorbed dose between the IPEMB protocol and the HPA/IPSM protocol vary from +0.5% to +1.6% at the depth of maximum dose. In addition the IPEMB measured doses are 0.2% larger than those measured following the IAEA code of practice. It may also be stated that the IPEMB measured doses at the depth of maximum dose are up to 1.5%, but generally less than 1.0%, lower than those measured by the AAPM protocol.

摘要

自1997年1月1日起,IPEMB最新的电子剂量测定实用规程成为英国电子束剂量测定的推荐方案,取代了先前的健康保护局实用规程及其IPSM附录。美国医学物理学家协会(AAPM)和国际原子能机构(IAEA)最近也针对在高能电子束中使用平行板电离室制定了电子束剂量测定的新建议。在此背景下,从将野外仪器电离室与二级标准进行比对,到确定电子束中参考位置处的吸收剂量,均遵循了这些实用规程中各自推荐的程序。已针对一系列电子束能量(从标称5兆电子伏到17兆电子伏)以及四种不同类型的野外仪器电离室测定了吸收剂量:一个NE2571石墨壁圆柱形电离室;一个NACP平行板电离室;一个马库斯平行板电离室;以及一个鲁斯平行板电离室。在最大剂量深度处,IPEMB规程与健康保护局/IPSM规程在吸收剂量测定上的差异在+0.5%至+1.6%之间。此外,IPEMB测量的剂量比按照国际原子能机构实用规程测量的剂量大0.2%。还可以指出,在最大剂量深度处,IPEMB测量的剂量比按照美国医学物理学家协会规程测量的剂量低达1.5%,但通常低于1.0%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验