Nisbet A, Thwaites D I
Department of Medical Physics and Medical Engineering, University of Edinburgh, Western General Hospital, UK.
Phys Med Biol. 1998 Feb;43(2):435-43. doi: 10.1088/0031-9155/43/2/016.
Polarity and ion recombination correction factors for the NACP (type-02) design parallel-plate ionization chamber employed in a recent UK national electron beam dosimetry intercomparison are derived over the full range of energies and measurement conditions encountered. In addition, these effects have been studied for a further four NACP chambers, a Markus parallel-plate chamber, a Roos parallel-plate chamber and a NE2571 graphite walled cylindrical ionization chamber.