Reynaert N, Verhaegen F, Thierens H
Department of Biomedical Physics, University of Gent, Belgium.
Phys Med Biol. 1998 Aug;43(8):2095-107. doi: 10.1088/0031-9155/43/8/008.
An ionometric calibration procedure for 192Ir PDR brachytherapy sources in terms of dose rate to water is presented. The calibration of the source is performed directly in a water phantom at short distances (1.0, 2.5 and 5.0 cm) using an NE2571 Farmer type ion chamber. To convert the measured air-kerma rate in water to dose rate to water a conversion factor (CF) was calculated by adapting the medium-energy x-ray dosimetry protocol for a point source geometry (diverging beam). The obtained CF was verified using two different methods. Firstly, the CF was calculated by Monte Carlo simulations, where the source-ionization chamber geometry was modelled accurately. In a second method, a combination of Monte Carlo simulations and measurements of the air-kerma rate in water (at 1.0, 2.5 and 5.0 cm distance) and in air (1 m distance) was used to determine the CF. The obtained CFs were also compared with conversion factors calculated with the adapted dosimetry protocol for high-energy photons introduced by Tölli. All calculations were done for a Gammamed PDR 192Ir source-NE2571 chamber geometry. The conversion factors obtained with the four different methods agree to within 1% at the three distances of interest. We obtained the following values (medium-energy x-ray protocol): CF(1 cm) = 1.458; CF(2.5 cm) = 1.162; CF(5.0 cm) = 1.112 (1 sigma = 0.7% for the three distances of interest). The obtained results were checked with TLD measurements. The values of the specific dose rate constant and the radial dose function calculated in this work are in accordance with the literature data.
本文介绍了一种针对192Ir PDR近距离放射治疗源的剂量率到水的离子测量校准程序。使用NE2571 Farmer型电离室在短距离(1.0、2.5和5.0厘米)的水模体中直接对源进行校准。为了将在水中测量的空气比释动能率转换为水的剂量率,通过调整适用于点源几何形状(发散束)的中能x射线剂量测定协议来计算转换因子(CF)。使用两种不同方法对获得的CF进行了验证。首先,通过蒙特卡罗模拟计算CF,其中精确模拟了源 - 电离室几何形状。在第二种方法中,结合蒙特卡罗模拟以及在水中(1.0、2.5和5.0厘米距离)和空气中(1米距离)的空气比释动能率测量来确定CF。还将获得的CF与Tölli引入的适用于高能光子的剂量测定协议计算的转换因子进行了比较。所有计算均针对Gammamed PDR 192Ir源 - NE2571室几何形状进行。在三个感兴趣的距离处,用四种不同方法获得的转换因子在1%以内相符。我们得到以下值(中能x射线协议):CF(1厘米)=1.458;CF(2.5厘米)=1.162;CF(5.0厘米)=1.112(对于三个感兴趣的距离,1σ = 0.7%)。用热释光剂量计测量对获得的结果进行了检验。在本工作中计算的比释动能率常数和径向剂量函数值与文献数据一致。