New polyether sulfone dialyzers attenuate passage of cytokine-inducing substances from pseudomonas aeruginosa contaminated dialysate.

作者信息

Jaber B L, Gonski J A, Cendoroglo M, Balakrishnan V S, Razeghi P, Dinarello C A, Pereira B J

机构信息

Division of Nephrology, Department of Medicine, New England Medical Center Hospitals, Boston, MA 02111, USA.

出版信息

Blood Purif. 1998;16(4):210-9. doi: 10.1159/000014336.

Abstract

The use of bicarbonate dialysate and high-flux and reprocessed dialyzers has raised concerns about the reverse transfer of dialysate contaminants into the blood compartment. This in vitro study was performed to investigate the reverse transfer of soluble Pseudomonas aeruginosa bacterial products across a polyether sulfone (PES), a newly developed synthetic polymer dialyzer. In vitro dialysis was carried out at 37 degreesC in a closed countercurrent recirculating loop dialysis circuit with a new PES dialyzer. An equal mixture of heparinized whole blood (from healthy volunteers) with pyrogen-free tissue culture medium was circulated in the blood compartment, and bicarbonate dialysate was circulated in the dialysate compartment. After 15 min of dialysis, the dialysate was challenged sequentially with 10(-4), 10(-3), and 10(-2) dilutions of a P. aeruginosa culture supernatant. 1-ml samples were drawn from the blood compartment 5 and 15 min after each challenge and incubated upright at 37 degrees C. At the end of 24 h, Triton X-100 was added, in order to extract total interleukin (IL) 6 and IL-8 production by the whole-blood mixture. These cytokines were measured by electrochemiluminescence assays. At dilutions of 10(-4) and 10(-3), the reverse transfer of soluble bacterial products across the dialyzer was negligible. Five and 15 min after contaminating the dialysate with the highest concentration (10(-2) dilution), the increase in IL-6 production was 239 +/- 170% (p = 0.06) and 886 +/- 444% (p = 0.02), respectively. However, comparing the IL-6-inducing potency of the 10(-2) bacterial supernatant dilution to the spontaneous IL-6 production in the blood compartment during dialysis with the same dilution of dialysate contaminant, there was a dramatic reduction in IL-6 production by 94 and 89% at 5 and 15 min, respectively. Similarly, 5 and 15 min after contaminating the dialysate with the 10(-2) dilution, the increase in IL-8 production was 357 +/- 147% (p = 0.07) and 630 +/- 229% (p = 0.04), respectively. However, comparing the IL-8-inducing potency of the 10(-2) bacterial supernatant dilution to the spontaneous IL-8 production in the blood compartment during dialysis with the same dilution of dialysate contaminant, there was a dramatic reduction in IL-8 production by 93 and 92% at 5 and 15 min, respectively. These results demonstrate that PES dialyzers markedly attenuate passage of cytokine-inducing substances from contaminated dialysate, using a method that detects the entire cytokine synthetic output in the blood compartment.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验