Suppr超能文献

Validity and reliability of measurements obtained with an "activity monitor" in people with and without a transtibial amputation.

作者信息

Bussmann H B, Reuvekamp P J, Veltink P H, Martens W L, Stam H J

机构信息

Institute of Rehabilitation Medicine, Faculty of Medicine and Health Sciences, Erasmus University Rotterdam, The Netherlands.

出版信息

Phys Ther. 1998 Sep;78(9):989-98. doi: 10.1093/ptj/78.9.989.

Abstract

BACKGROUND AND PURPOSE

In this study, the validity and reliability of measurements obtained with an "Activity Monitor" (AM) were examined. The instrument is designed to monitor ambulatory activity by use of accelerometer signals, and it detects several activities associated with mobility (standing, sitting, lying, transitions, movement-related activities).

SUBJECTS

Four men with a transtibial amputation and 4 men without a transtibial amputation participated.

METHODS

The subjects performed normal daily activities, during which accelerations were measured and videotape recording were made (reference method). Validity was assessed by calculating agreement scores between the AM output and the videotape recordings and by comparing the number of transitions and the duration of activities determined by both methods.

RESULTS

The overall agreement between the AM output and the videotape recordings was 90%. Other agreement scores, in addition to the determination of the number of transitions and the duration of activities, were generally within a range of error of 0% to 10%.

CONCLUSION AND DISCUSSION

The reliability and validity of the AM measurements appeared to be good, which supports its potential use in rehabilitation and physical therapy. [Bussmann HBJ, Reuvekamp PJ, Veltink PH, et al. Validity and reliability of measurements obtained with an "Activity Monitor" in people with and without a transtibial amputation.