Alfke H, Ishaque N, Froelich J J, Klose K J
Abteilung Strahlendiagnostik, Philipps Universität, Marburg.
Radiologe. 1998 Jul;38(7):591-6. doi: 10.1007/s001170050397.
To evaluate the image quality of magnetic resonance venography (MRV) of pelvic and abdominal veins.
A retrospective analysis of all MRV between 1993 and 1996 was conducted. A 2D-FLASH-TOF without breath hold and with arterial presaturation was used. All venous segments have been assessed for the quality of visualisation and an overall image quality was determined. 43% of our patients had malignant disease and phlebothrombosis had already been diagnosed in 64%.
126 examinations and 1696 venous segments were analysed. The overall quality of the examination was good, however one examination was non diagnostic because of motion artefacts. Vessels along the vertical axis were good visualised in over 90%. Most important reasons for insufficient visualisation were breathing, motion and metal artefacts as well as compression and displacement of veins due to tumour disease.
Magnetic resonance venography without breath hold TOF technique is practical and robust for routine clinical applications.
评估盆腔和腹部静脉磁共振静脉血管造影(MRV)的图像质量。
对1993年至1996年间所有的MRV进行回顾性分析。采用二维快速扰相梯度回波(2D-FLASH-TOF)序列,无需屏气并使用动脉预饱和技术。对所有静脉节段的可视化质量进行评估,并确定整体图像质量。43%的患者患有恶性疾病,64%的患者已被诊断为静脉血栓形成。
分析了126例检查和1696个静脉节段。检查的整体质量良好,但有一次检查因运动伪影而无法诊断。垂直轴上的血管可视化率超过90%。可视化不足的最重要原因是呼吸、运动和金属伪影,以及肿瘤疾病导致的静脉受压和移位。
无需屏气的TOF技术磁共振静脉血管造影在常规临床应用中实用且可靠。