To measure whole-body dose in tomotherapy of the head and neck region resulting from internal patient scatter and linear accelerator leakage.
Treatments are performed using a commercial computer-controlled intensity modulated radiation therapy planning and delivery system (Peacock, NOMOS Corp.) and a 6-MV linear accelerator (Clinac 6/100, Varian Corp.). The patient dose outside the treatment field is measured in a water-equivalent phantom using thermoluminescent dosimetry. The whole-body dose components from internal scatter and leakage are separately determined. The use of fixed-portal leakage and scattered radiation measurements to estimate the whole-body dose from tomotherapy is evaluated.
The internally scattered dose is significant near the target, but becomes negligible relative to the leakage dose beyond 15 cm from the target. Dose at 10 cm from the target volume, due to internal scatter and leakage, is approximately 2.5% of the total target dose, reducing to 0.5% at 30 cm. The measured dose is relatively uniform throughout the phantom.
The whole-body dose equivalent from a tomotherapy treatment is greater than that from conventional radiation therapy. Further studies are required to assess the trade-off between improved dose distribution conformality and a possible slight increase in radiation-induced fatal malignancies. The accuracy of using fixed-portal leakage and scattered dose measurements to estimate the whole-body dose from tomotherapy treatments is adequate, if the appropriate fixed-portal field size equivalent is used.
测量头颈部区域断层放射治疗中因患者体内散射和直线加速器泄漏所导致的全身剂量。
使用商用计算机控制的调强放射治疗计划与投送系统(孔雀系统,NOMOS公司)以及一台6兆伏直线加速器(Clinac 6/100,瓦里安公司)进行治疗。在水等效体模中使用热释光剂量测定法测量治疗野之外的患者剂量。分别确定来自体内散射和泄漏的全身剂量分量。评估使用固定射野泄漏和散射辐射测量来估算断层放射治疗全身剂量的情况。
在靶区附近,体内散射剂量显著,但在距靶区15厘米以外相对于泄漏剂量可忽略不计。距靶区体积10厘米处,由于体内散射和泄漏产生的剂量约为总靶区剂量的2.5%,在30厘米处降至0.5%。在整个体模中测量的剂量相对均匀。
断层放射治疗的全身剂量当量大于传统放射治疗。需要进一步研究来评估在改善剂量分布适形性与辐射诱发致命恶性肿瘤可能略有增加之间的权衡。如果使用适当的固定射野等效野大小,使用固定射野泄漏和散射剂量测量来估算断层放射治疗全身剂量的准确性是足够的。