Le Du I, Faucheux M, Lurton Y, Basle B, Mallédant Y
Centrale d'achat du matériel stérile et du pansement, CHU Rennes, France.
Ann Fr Anesth Reanim. 1998;17(1):32-40. doi: 10.1016/s0750-7658(97)80179-7.
To compare the time course of endotracheal tube cuff pressures in presence of nitrous oxide (N2O), obtained in a tracheal model with those measured during clinical anaesthesia.
Experimental and clinical prospective study.
Twelve brands of low-pressure tracheal tubes.
The pressure changes in the cuffs were measured over a three-hour-period in presence of a N2O (50 vol%)/O2 (50 vol%) mixture and mechanical ventilation, the tube being inserted either in a tracheal model or in the trachea of patients during general anaesthesia.
The results obtained in vitro were correlated with those measured in the patients. Therefore the tracheal model is a helpful guide for the choice of endotracheal tubes.