Beam characteristics of a retrofitted double-focused multileaf collimator.

作者信息

Das I J, Desobry G E, McNeeley S W, Cheng E C, Schultheiss T E

机构信息

Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, Pennsylvania 19111, USA.

出版信息

Med Phys. 1998 Sep;25(9):1676-84. doi: 10.1118/1.598348.

Abstract

Multileaf collimators (MLCs) are generally believed to be convenient and cost-effective tools for intensity modulation and conformal therapy. They are becoming a standard feature on new accelerators; however, the older units can be retrofitted with modern MLCs. Before such a unit can be clinically used, the beam characteristics must be verified. In this study the beam characteristics of a Siemens double-focused MLC retrofitted to an MD2 linear accelerator are presented. The head leakage along with inter- and intra-leaf radiation transmission were measured using film. The collimator (Sc), phantom (Sp), total (Scp) scatter factors, central axis depth dose, beam profiles for off-axis ratios, penumbra, and surface dose were evaluated for square, rectangular, and irregularly shaped fields. The maximum head leakage was estimated to be < 0.05% in any plane at a distance of 1 m and maximum transmission through the MLC leaves was estimated to be < 1.4% and < 1.1% for the 10 MV and 6 MV beams, respectively. The maximum differences between pre- and post-MLC installation data for the Sc and Scp were < or = 0.7% and < or = 1.4%, respectively. Similarly, the percent depth dose data for all fields and both beam energies were within 1.5% of the original data. The beam profiles measured at various depths were also in agreement with those of the pre-MLC installation data. The measured beam penumbra (20%-80%) showed a range of 7.8 mm-11.0 mm for the 6 MV and 8.4 mm-11.1 mm for the 10 MV beams from smallest to largest fields. These ranges differ by less than a millimeter from those of the old data. The surface dose measurements were slightly lower than the conventional jaw values suggesting that MLC does not produce significant electron contamination. It is concluded that the retrofitted MLC maintains the integrity of the original beam and may provide a cost-effective conformal therapy.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验