[头颈部位置改变对气管导管套囊的影响]

[Effects of changes in head and neck position on a tracheal tube cuff].

作者信息

Inoue S, Takauchi Y, Kuro M, Ninaga H

机构信息

Department of Anesthesiology, National Cardiovascular Center, Suita.

出版信息

Masui. 1998 Sep;47(9):1069-72.

PMID:9785780
Abstract

Thirteen patients were intubated with cuffed reinforced spiral tracheal tubes. Intracuff pressure and volume were measured as the position of the head and neck was altered. No significant changes in intracuff pressure and volume were observed with lateral rotation of the head. A significant increase in intracuff pressure was observed and excessive pressure (> 25 cm H2O) developed in 8 patients with head and neck flexion. A significant decrease of intracuff pressure and air-leak were observed in 9 patients with extension. Reinflated intracuff volume decreased and reinflated intracuff pressure increased significantly, and residual excessive pressure was observed in 4 patients with flexion. Both reinflated intracuff volume and pressure increased significantly, and excessive pressure was observed in 8 patients and residual air-leak developed in a patient in spite of excessive pressure with extension. The authors speculate that endotracheal tube movement by changes in head and neck position has effects on intracuff pressure and volume.

摘要

13例患者使用带套囊的加强型螺旋气管导管进行插管。随着头颈部位置的改变,测量套囊内压力和容积。头部侧旋时,套囊内压力和容积未观察到显著变化。8例头颈部屈曲患者观察到套囊内压力显著升高并出现过高压力(>25 cm H₂O)。9例头颈部伸展患者观察到套囊内压力显著降低及漏气。4例屈曲患者再充气后套囊容积减小,套囊压力显著升高,并观察到残余过高压力。头颈部伸展时,尽管存在过高压力,但8例患者再充气后套囊容积和压力均显著增加,1例患者出现残余漏气。作者推测,头颈部位置改变导致气管导管移动会对套囊内压力和容积产生影响。