Loos R, Niessner R
Institute of Hydrochemistry, Technical University of Munich, Germany.
J Chromatogr A. 1998 Oct 2;822(2):291-303. doi: 10.1016/s0021-9673(98)00620-7.
The separation of 14 different aromatic sulfonates of environmental concern by capillary (zone) electrophoresis (CZE) is presented. A new off-line solid-phase extraction (SPE) enrichment procedure, that is compatible with CE analysis, was developed, using the styrene-divinylbenzene adsorbent LiChrolut EN. The combined method of SPE and CE allows the determination of aromatic sulfonates in water samples in the low microgram/l range. Separations are performed with a simple sodium borate buffer at pH 9.3. Analytes are detected by UV absorbance and fluorescence emission with a Xe-lamp excitation source, and both principles are compared. The recoveries for most of the sulfonates are > 70% for the extraction from spiked tap and river water. The average method precision is < 20% for replicate analyses. Very hydrophilic sulfonates cannot be extracted by this method. The detection limit of the combined method of SPE enrichment and CE analysis is approximately 0.1 microgram/l for 200-ml water samples. The performance of the method was checked with the analysis of river and contaminated seepage water.