Lindberg P, Roeraade J
Royal Institute of Technology, Department of Analytical Chemistry, Stockholm, Sweden.
Electrophoresis. 1998 Oct;19(14):2445-6. doi: 10.1002/elps.1150191415.
A novel method for the manufacture of detection windows in polyimide-coated fused-silica capillary columns is described. Concentrated fuming nitric acid at room temperature is utilized for stripping of the coating from a defined section of the capillary. The technique is particularly suitable for columns that have a heat sensitive coating on the inner wall or for columns that contain an immobilized hydrogel sieving-matrix. A column holder for the simultaneous preparation of detection windows in an array of more than 100 capillaries is also reported.
描述了一种在聚酰亚胺涂层熔融石英毛细管柱中制造检测窗口的新方法。室温下的浓发烟硝酸用于从毛细管的特定部分剥离涂层。该技术特别适用于内壁有热敏涂层的柱子或含有固定化水凝胶筛分基质的柱子。还报道了一种用于同时在100多个毛细管阵列中制备检测窗口的柱架。