Price K W, Nath R, Holeman G R
Med Phys. 1978 Jul-Aug;5(4):285-9. doi: 10.1118/1.594479.
High-energy x-ray radiotherapy machines generate neutrons by photonuclear reactions in the target and the treatment head and expose the patient to a neutron flux. In order to evaluate the neutron exposure quantitatively, fast and thermal neutron profiles for 25-MV x-ray beams of the Sagittaire accelerator have been measured. An activation technique, using the reactions 31P(n, gamma)32P (thermal neutrons) and 31P(n, p)31Si (fast neutrons, E greater than 0.7 MeV), has been developed to measure fast- and thermal-neutron fluxes in an intense high-energy photon flux. The sensitivity of this activation detector to high-energy photons, which has plagued many previous neutron measurements, was carefully measured and found to be less than 4%. Neutron fluxes for various photon field sizes ranging from 5 X 5 cm to 30 X 30 cm have been measured. The fast-neutron profiles were observed to have rounded edges and the thermal fluxes were found to be relatively uniform. In the central part of the x-ray beam, the ratio of neutron dose equivalent to photon absorbed dose was found to be between 0.2% and 0.5%. Outside of the photon field, the ratio of neutron dose equivalent to the central-axis photon absorbed dose was 0.12%.
高能X射线放射治疗机通过靶区和治疗头中的光核反应产生中子,并使患者暴露于中子通量中。为了定量评估中子暴露情况,已对射手座加速器25兆伏X射线束的快中子和热中子分布进行了测量。已开发出一种活化技术,利用31P(n,γ)32P反应(热中子)和31P(n,p)31Si反应(快中子,能量大于0.7兆电子伏)来测量强高能光子通量中的快中子和热中子通量。仔细测量了这种活化探测器对高能光子的灵敏度,该灵敏度在之前许多中子测量中一直是个问题,结果发现其小于4%。已测量了从5×5厘米到30×30厘米各种光子场大小的中子通量。观察到快中子分布的边缘呈圆形,热通量相对均匀。在X射线束的中心部分,发现中子剂量当量与光子吸收剂量的比值在0.2%至0.5%之间。在光子场之外,中子剂量当量与中心轴光子吸收剂量的比值为0.12%。