Dose levels due to neutrons in the vicinity of high-energy medical accelerators.

作者信息

McGinley P H, Wood M, Mills M, Rodriguez R

出版信息

Med Phys. 1976 Nov-Dec;3(6):397-402. doi: 10.1118/1.594256.

Abstract

High-energy photons are generated for use in radiation therapy by the deceleration of electrons in metal targets. Fast neutrons are also generated as a result of (gamma, n) and (e, e'n) interactions in the target, beam compensator filter, and collimator material. In this work the absorbed dose due to neutrons was measured at the center of a 10 X 10-cm photon beam and 5 cm outside of the beam edge for a number of treatment units (Varian Clinac-18, Allis-Chalmers 25-MeV betatron, and Brown Boveri 45-MeV betatron). Dose levels due to slow and fast neutrons were also established outside of the treatment rooms and a Bonner sphere neutron spectrometer system was employed to determine the neutron energy spectrum due to stray neutron radiation at each accelerator. For the linac it was found that the neutron dose at the beam center was 0.0039% of the photon dose and values of 0.04% and 0.053% were observed for the Allis-Chalmers betatron and the Brown Boveri betatron. Dose-equivalent rates in the range 0.3-22.5 mrem were measured for points outside the treatment rooms when the accelerators were operated at a photon dose rate of 100 rad/min at the treatment position.

摘要

通过电子在金属靶中的减速来产生用于放射治疗的高能光子。在靶、射束补偿滤波器和准直器材料中,由于(γ,n)和(e,e'n)相互作用也会产生快中子。在这项工作中,对多个治疗单元(瓦里安Clinac - 18、阿利斯 - 查默斯25兆电子伏特电子感应加速器和布朗·博韦里45兆电子伏特电子感应加速器),测量了10×10平方厘米光子束中心以及束边缘外5厘米处的中子吸收剂量。还在治疗室外部确定了慢中子和快中子的剂量水平,并使用邦纳球中子谱仪系统来确定每个加速器处杂散中子辐射产生的中子能谱。对于直线加速器,发现在束中心的中子剂量是光子剂量的0.0039%,而对于阿利斯 - 查默斯电子感应加速器和布朗·博韦里电子感应加速器,观察到的值分别为0.04%和0.053%。当加速器在治疗位置以100拉德/分钟的光子剂量率运行时,在治疗室外部的点测量到的剂量当量率范围为0.3 - 22.5毫雷姆。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验