de Kanel J, Vickery W E, Diamond F X
National Medical Services, Inc., Willow Grove, Pennsylvania, USA.
J Am Soc Mass Spectrom. 1998 Mar;9(3):255-7. doi: 10.1016/S1044-0305(97)00289-4.
A method is described for the simultaneous analysis of 14 non-steroidal anti-inflammatory drugs (NSAIDS) in human serum using negative electrospray ionization-tandem mass spectrometry (ESI-MS/MS). After addition of internal standard and protein precipitation using acetonitrile, samples were transferred to autosampler vials for direct analysis without chromatography. Injection of an air bubble with the sample and a multiple reaction monitoring (MRM) method using argon collision-induced dissociation (CID) of analyte (M-H)- ions permitted integration of the product ion peak areas to produce reproducible quantitative data over the range of concentrations expected in serum during routine use of these drugs. The method permitted the analysis of 30 samples per hour. Two hundred fifty consecutive analyses did not adversely affect instrument sensitivity.
本文描述了一种使用负电喷雾电离串联质谱法(ESI-MS/MS)同时分析人血清中14种非甾体抗炎药(NSAIDs)的方法。加入内标并用乙腈进行蛋白沉淀后,将样品转移至自动进样瓶中,无需色谱分离即可直接分析。将气泡与样品一起进样,并采用多反应监测(MRM)方法,利用氩气碰撞诱导解离(CID)分析物(M-H)-离子,通过对产物离子峰面积进行积分,可在这些药物常规使用期间血清预期浓度范围内产生可重复的定量数据。该方法每小时可分析30个样品。连续250次分析对仪器灵敏度没有不利影响。