VITEK 45 wells Card was evaluated. Recent clinical isolates of gram-positive cocci (107) and gram-negative bacilli (179) were antimicrobial susceptibility testing by VITEK system in comparison with MIC-2000 system. The antimicrobial susceptibility results were available within of 8 h for gram-positive cocci without a part of MRSA. Enterobac-teriaceae were available within of 8 h and Pseudomonas aeruginosa were from 8 to 11 h. The antimicrobial susceptibility data had disagreement of 60% with EM, 46% with CLDM for Enterococcus faecium, 50% with CMZ for Citrobacter freundii, and 71% with CTM for Morganella morganii, whereas it was satisfactory agreement of 97.5% for gram-positive cocci and 98.2% for gram-negative bacilli. In using this procedure it was possible to provide accurate and rapid results of anti-microbial susceptibility tests for all organisms.

摘要

对VITEK 45孔卡片进行了评估。采用VITEK系统对近期分离出的革兰氏阳性球菌(107株)和革兰氏阴性杆菌(179株)进行抗菌药物敏感性试验,并与MIC - 2000系统进行比较。对于不包括耐甲氧西林金黄色葡萄球菌(MRSA)的革兰氏阳性球菌,8小时内可获得抗菌药物敏感性结果。肠杆菌科细菌在8小时内可获得结果,铜绿假单胞菌在8至11小时内可获得结果。粪肠球菌的抗菌药物敏感性数据与纸片扩散法(EM)的不一致率为60%,与氯碳头孢(CLDM)的不一致率为46%,弗氏柠檬酸杆菌与头孢孟多(CMZ)的不一致率为50%,摩根摩根菌与头孢噻肟(CTM)的不一致率为71%,而革兰氏阳性球菌的一致性令人满意,为97.5%,革兰氏阴性杆菌的一致性为98.2%。使用该方法可以为所有微生物提供准确、快速的抗菌药物敏感性试验结果。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验