Richard K E, Block F R, Weiser R R
Neurosurgical Hospital of Heinrich-Heine University, Düsseldorf, Germany.
Neurol Res. 1999 Jan;21(1):117-20. doi: 10.1080/01616412.1999.11740906.
The telemetric shunt-integrated sensors available up to now do not deliver negative CSF pressure values. Therefore, a new sensor was designed which is incorporated into the shunt line proximally to the valve. The monitor calculates the pressure values in relation to the atmospheric pressure obtained by a built-in barometer. The sensor covers a pressure range of -100 cmH2O to 200 cmH2O with a resolution of +/- 1 cmH2O. First follow-up experiences are reported with seven patients over 14-17 months. Depending on the patient's body position telemetrically measured CSF-pressures varied between -20 cmH2O in erect and 15 cmH2O in supine position. In a patient with intermittent shunt dysfunction and stiff ventricular walls telemetrically measured CSF values corresponded earlier and were more sensitive with the clinical symptomatology than with the CT scan. So far, recognizable advantages of the telesensor are an improved assessment of shunt dysfunction and marginal CSF pressure increases, with practicability and simplicity of measurements in outpatients and a reduction of expensive CT or MRI controls.
目前可用的遥测分流集成传感器无法提供脑脊液负压值。因此,设计了一种新的传感器,它被安装在靠近阀门的分流管线上。该监测器根据内置气压计获得的大气压力来计算压力值。该传感器的压力范围为-100 cmH₂O至200 cmH₂O,分辨率为±1 cmH₂O。报告了7例患者在14至17个月内的首次随访经验。根据患者的体位,遥测测量的脑脊液压力在直立位时为-20 cmH₂O,仰卧位时为15 cmH₂O之间变化。在一名患有间歇性分流功能障碍和脑室壁僵硬的患者中,遥测测量的脑脊液值比CT扫描更早且更敏感地与临床症状相符。到目前为止,遥测传感器的明显优点是改善了对分流功能障碍和脑脊液压力轻微升高的评估,在门诊患者中测量具有实用性和简便性,并减少了昂贵的CT或MRI检查。