Tapiovaara M J, Sandborg M, Dance D R
STUK-Radiation and Nuclear Safety Authority, Helsinki, Finland.
Phys Med Biol. 1999 Feb;44(2):537-59. doi: 10.1088/0031-9155/44/2/018.
A Monte Carlo computational model of a fluoroscopic imaging chain was used for deriving optimal technique factors for paediatric fluoroscopy. The optimal technique was defined as the one that minimizes the absorbed dose (or dose rate) in the patient with a constraint of constant image quality. Image quality was assessed for the task of detecting a detail in the image of a patient-simulating phantom, and was expressed in terms of the ideal observer's signal-to-noise ratio (SNR) for static images and in terms of the accumulating rate of the square of SNR for dynamic imaging. The entrance air kerma (or air kerma rate) and the mean absorbed dose (or dose rate) in the phantom quantified radiation detriment. The calculations were made for homogeneous phantoms simulating newborn, 3-, 10- and 15-year-old patients, barium and iodine contrast material details, several x-ray spectra, and for imaging with or without an antiscatter grid. The image receptor was modelled as a CsI x-ray image intensifier (XRII). For the task of detecting low- or moderate-contrast iodine details, the optimal spectrum can be obtained by using an x-ray tube potential near 50 kV and filtering the x-ray beam heavily. The optimal tube potential is near 60 kV for low- or moderate-contrast barium details, and 80-100 kV for high-contrast details. The low-potential spectra above require a high tube load, but this should be acceptable in paediatric fluoroscopy. A reasonable choice of filtration is the use of an additional 0.25 mm Cu, or a suitable K-edge filter. No increase in the optimal tube potential was found as phantom thickness increased. With the constraint of constant low-contrast detail detectability, the mean absorbed doses obtained with the above spectra are approximately 50% lower than those obtained with the reference conditions of 70 kV and 2.7 mm Al filter. For the smallest patient and x-ray field size, not using a grid was slightly more dose-efficient than using a grid, but when the patient size and field size were increased a fibre interspaced grid resulted in lower doses than imaging without a grid. For a 15-year-old patient the mean absorbed doses were up to 40% lower with this grid than without the grid.
使用荧光透视成像链的蒙特卡罗计算模型来推导儿科荧光透视的最佳技术参数。最佳技术被定义为在图像质量恒定的约束下使患者吸收剂量(或剂量率)最小化的技术。针对检测患者模拟体模图像中的细节任务评估图像质量,对于静态图像,用理想观察者的信噪比(SNR)表示,对于动态成像,用SNR平方的累积率表示。体模中的入射空气比释动能(或空气比释动能率)和平均吸收剂量(或剂量率)量化了辐射危害。针对模拟新生儿、3岁、10岁和15岁患者的均匀体模、钡剂和碘剂造影剂细节、几种X射线光谱以及有无反散射栅成像进行了计算。图像接收器被建模为碘化铯X射线影像增强器(XRII)。对于检测低对比度或中等对比度碘细节的任务,通过使用接近50 kV的X射线管电压并对X射线束进行大量过滤可获得最佳光谱。对于低对比度或中等对比度钡细节,最佳管电压接近60 kV,对于高对比度细节,最佳管电压为80 - 100 kV。上述低电压光谱需要高管负载,但这在儿科荧光透视中应是可接受的。合理的过滤选择是使用额外的0.25 mm铜或合适的K边滤波器。未发现随着体模厚度增加最佳管电压有所增加。在低对比度细节可检测性恒定的约束下,使用上述光谱获得的平均吸收剂量比在70 kV和2.7 mm铝滤波器的参考条件下获得的剂量低约50%。对于最小的患者和X射线视野尺寸,不使用栅比使用栅在剂量效率上略高,但当患者尺寸和视野尺寸增加时,纤维间隔栅导致的剂量低于无栅成像。对于15岁患者,使用这种栅时的平均吸收剂量比不使用栅时低多达40%。