Hilfiker P R, Weishaupt D, Schmid M, Dubno B, Hodler J, Debatin J F

机构信息

Institute of Diagnostic Radiology and Radiology, University Hospital Zurich, Switzerland.

出版信息

Eur Radiol. 1999;9(2):201-4. doi: 10.1007/s003300050655.

Abstract

The purpose of this study was to evaluate interactive MR-guided joint puncture with intra-articular application of contrast agent. MR-guided arthrography of the shoulder joint was successfully performed in three patients using an interactive guidance system implemented in an open-configuration MR system. Visualization of the needle pathway and contrast inflow was comparable to that with conventional X-ray fluoroscopy. The position of the intra-articular needle tip was accurately confirmed and subsequent MR arthrography was diagnostic in all cases.

摘要

本研究的目的是评估在关节内应用造影剂的交互式磁共振引导下关节穿刺。使用开放式磁共振系统中实施的交互式引导系统,成功地对三名患者进行了肩关节磁共振引导下关节造影。针道和造影剂流入的可视化与传统X线透视相当。关节内针尖的位置得到准确确认,所有病例随后的磁共振关节造影均具有诊断价值。