Ter-Pogossian M M, Mullani N A, Hood J T, Higgins C S, Ficke D C
J Comput Assist Tomogr. 1978 Nov;2(5):539-44. doi: 10.1097/00004728-197811000-00002.
Imaging of the brain by positron emission tomography can be optimized for sensitivity by dedicating the design of the tomograph to this application. We have designed a multislice positron emission tomograph (PETT V) for imaging the human brain and the whole body of small experimental animals. The detector system of PETT V consists of a circular array of 48 NaI(Tl) scintillation detectors, each fitted with two photomultiplier tubes, with one dimensional positioning capability. Suitable sampling is achieved by rotation of the circular array of detectors and by a wobbling motion of the detector circle. The proposed system is capable of providing seven slices simultaneously, with a spatial resolution in the plane of the slice from 7 to 15 mm and with slice thicknesses of 7 and 14 mm. The minimum scanning time is 1 sec. The estimated overall sensitivity of PETT V is 350,000 counts/sec/mCi in a 20 cm diameter phantom for a resolution of approximately 1.5 x 1.5 cm. The system is under construction.
通过将断层扫描仪的设计专门用于此应用,可以优化正电子发射断层扫描对大脑的成像灵敏度。我们设计了一种多层正电子发射断层扫描仪(PETT V),用于对人类大脑和小型实验动物的全身进行成像。PETT V的探测器系统由48个NaI(Tl)闪烁探测器的圆形阵列组成,每个探测器配备两个光电倍增管,具有一维定位能力。通过探测器圆形阵列的旋转和探测器圆的摆动运动实现合适的采样。所提出的系统能够同时提供七个切片,切片平面内的空间分辨率为7至15毫米,切片厚度为7和14毫米。最短扫描时间为1秒。在直径20厘米的体模中,对于分辨率约为1.5×1.5厘米的情况,PETT V的估计总体灵敏度为350,000计数/秒/毫居里。该系统正在建设中。