In ion chromatography, samples of very different ammonium-to-sodium concentration ratios are difficult to quantify since these two cations have similar selectivities for stationary phases containing commonly used sulfonate or carboxylate cation-exchange functional groups. The IonPac CS15 cation-exchange column, with carboxylate and phosphonate functional groups as well as a crown ether group, was developed to address this limitation. Selectivity for the common inorganic cations on this column is different from that of conventional cation-exchange columns in that the separation between sodium and ammonium ions has been greatly increased, allowing for determinations of low levels of one in the presence of high levels of the other with an isocratic eluent. For larger than 4000:1 sodium-to-ammonium concentration ratios, an eluent step change or gradient elution is needed. For moderate ratios, combinations of this column with a carboxylate column, containing no crown ether group, can be used at room temperature with an isocratic eluent containing no organic solvent.
在离子色谱法中,铵钠浓度比差异很大的样品难以定量,因为这两种阳离子对含有常用磺酸盐或羧酸盐阳离子交换官能团的固定相具有相似的选择性。开发了具有羧酸盐、膦酸盐官能团以及冠醚基团的IonPac CS15阳离子交换柱来解决这一限制。该柱对常见无机阳离子的选择性与传统阳离子交换柱不同,因为钠和铵离子之间的分离度大大提高,使用等度洗脱液时,在一种离子含量高的情况下也能测定另一种离子的低含量。对于钠铵浓度比大于4000:1的情况,需要进行洗脱液阶跃变化或梯度洗脱。对于中等比例的情况,该柱与不含冠醚基团的羧酸盐柱组合,可在室温下使用不含有机溶剂的等度洗脱液。